Owner manual

CF 126 42 Nov 2001
Using pp="01" accesses the "CALIBRATION TYPE" menu item.
When writing this parameter:
n=1 means Shunt Calibration.
n=2 means 2-Point Known-Load Calibration.
n=3 means 3-Point Known-Load Calibration.
n=5 means 5-Point Known-Load Calibration.
Using pp="02" accesses the "AUX1 FUNCTION" menu item.
Using pp="03" accesses the "AUX2 FUNCTION" menu item.
When writing these parameters:
n=0 means the Auxiliary Function pin is disabled.
n=1 means track hold.
n=2 means peak & valley hold.
n=4 means peak & valley clear (edge triggered).
n=16 means activate Tare function.
n=32 means deactivate Tare function.
Example
Sending "#0001WPO216" will allow the assertion of channel 01 's Auxiliary 1 Function
pin to activate the Tare function. The instrument will reply with "OK".
Reply When writing: "OK" or "ERROR"
When reading: a numeric value according to the information above.
RQ/WQ Read/write Display Formatting
Purpose Reads or writes the operation of the channel's display formatting.
Usage
"#aaccRQ" to read, "#aaccWQn" to write.
# is the 'pound' or 'hash' character (ASCII decimal 35).
aa is the two-character instrument address.
cc is the two-character channel number
n is the argument defined below.
is the 'carriage return' character (ASCII decimal 13).
Argument The argument is created by adding together the values of the desired options as shown.
Digits
Value Decimal Point Value
5 (bipolar) 0. 0 0.
6 (Unipolar) 32. 1 1.
7 (Unipolar) 3104. 2 2.
6 3.
7 4.
8 5.
Count-By
Value
1 0.
2 152. Averaging
Value
5 280. OFF 0.
10 8. ON 64.
20 408.
100 16.
200 664.
Example It is desired to set channel 08's display to operate as shown:
Two digits after decimal point = 2
5 digit, bipolar display = 0
Count by one digit = 0
Enable display averaging = 64
Sending "#0008WQ66" will change display operation accordingly.
Reply When writing: "OK" or "ERROR".
When reading: an ASCll-floating-point value described above.
RR Read Version Info
Purpose Read version information found on the "VERSION INFO" menu item.