Manual

CF 66 23 Rev. C 2/05
Figure 9-2: Limit Operation when LIMIT.ENERGIZE = SIGNAL < SETPOINT
Figure 9-3: Limit Operation when LIMIT.ENERGIZE = SIGNAL INSIDE
Figure 9-3: Limit Operation when LIMIT.ENERGIZE = SIGNAL OUTSIDE
9.3.6 LIMIT.LATCHING Menu Item
This specifies whether to latch the activated limit so that only manually clearing the limit will deactivate it. A latched
limit can be manually cleared by pressing the [CLEAR] button, sending the “F8” serial communications command,
or by using the Function Input #2 (Clear Peak/Valley & Limits pin on the System Connector. See “Function Input
Pins” in Section 5.3.
The options for this menu item are:
“ON” means that once a signal activates the limit, the limit will remain activated until it is reset
“OFF” means that the limit will deactivate immediately when the signal crosses the return point.
9.3.7 LIMIT.CHANNEL Menu Item
This specifies the channel that is monitored by this limit.