DFI 1650 SIGNAL CONDITIONING, SELF-CALIBRATING INDICATOR USER’S GUIDE www.cooperinstruments.
CONTENTS 1.0 INTRODUCTION ................................................................................................................... 1 1.1 About This Manual ......................................................................................................................... 1 1.1.1 Scope......................................................................................................................................................... 1 1.1.2 Conventions......................................
4.4 DFI 1650 PT................................................................................................................................... 11 4.4.1 Differences.............................................................................................................................................. 11 4.4.2 External Arrangement............................................................................................................................ 11 4.4.3 Front Panel............................
8.2.1 UPPER CHANNEL Menu Item ............................................................................................................... 20 8.2.2 LOWER CHANNEL Menu Item .............................................................................................................. 20 8.2.3 LOWER MODE Menu Item ..................................................................................................................... 20 8.2.4 DISPLAY DISABLE Menu Item ...........................................
11.5.7 DAC SETUP Sub-Menu ........................................................................................................................ 42 11.5.8 DIAGNOSTICS Sub-Menu .................................................................................................................... 43 11.6 Electrical Null and Transducer Mounting ................................................................................ 43 11.6.1 Overview ....................................................................
14.0 DAC OUTPUT CHANNEL................................................................................................. 62 14.1 Features ...................................................................................................................................... 62 14.2 Wiring.......................................................................................................................................... 62 14.3 Setup Procedure ..............................................................
1.0 INTRODUCTION 1.1 About This Manual 1.1.1 Scope This manual will explain the setup, features and operation of Cooper’s digital force indicator instruments. This series includes the DFI 1650, 1650PT and SC3004. Further information about customer specific programming and setup will be explained on the System Calibration Sheet that is provided with every instrument. 1.1.2 Conventions This manual uses the following conventions to present information: [TEXT IN BRACKETS] The label of a front panel button.
Chapter 14, “DAC Output Channel”, explains the configuration and operation of additional digital-to-analog voltage or current outputs. Chapter 15, “Split Display Virtual Channel”, shows how you can display two values from any of the channels in the instrument at the same time. Chapter 16, “Mathematics Virtual Channel”, describes the flexibility of customer specific SensoCode programming.
• • • • • Strain-gage sensors, such as un-amplified pressure transducers and load cells Sensors with voltage outputs, such as transducers with the Option 2a, 2b, 2c, or 2t internal amplifier. Sensors with current outputs, such as transducers with the Option 2j, 2k or 2n internal amplifier AC-AC LVDT’s (Linear Variable Displacement Transducers) DC-DC LVDTs Output Channels Output channels are hardware circuit boards with a unique channel number.
additional “multiple-point calibration” can be purchased with the transducer that allows linearity correction information to be placed into its Signature Calibration module. 2.0 GETTING STARTED QUICKLY 2.1 Locate Required Parts and Information The following items are required to set up a DFI instrument with your transducer: • DFI instrument • Transducers that are to be connected to the instrument • For each transducer to be connected to the instrument, a connecting cable.
• Value type: The next character indicates the nature of the following value. A blank character indicates the tracking value. The instrument is “tracking” the signal, continuously updating the display in response to the signal from the transducer. • A ”▲“ character indicates the peak value (highest value seen since the [CLEAR] button was last pressed). • A “▼“ character indicates the valley value (lowest value seen since the [CLEAR] button was last pressed). • Data value: “00000.
3.3 RUN Mode After the INITIALIZE mode finishes, the instrument enters the RUN mode, its normal mode of operation. 3.3.1 Display DFI SC3004 and upper line of DFI 1550, DFI 1650, DFI 1650PT The display will show a channel number on the far left, followed by the channel’s operation messages. For example, a Strain Gage amplifier channel will use the format below: 1▲ 00000. PSIG ◊ where: • Channel number: “1” is the channel number • Value type: The next character indicates the nature of the following value.
On instruments with a dual-line display, the [CHANNEL] button cannot be used to change which channel the lower line is monitoring. That is selected with the “DISPLAY MENU->LOWER CHANNEL” menu item. If the [CHANNEL] button is held down for more than 3 seconds, the present channel’s shunt calibration value (if the channel has shunt calibration available) will be displayed. 3.3.5 [TARE] button To reset the channel’s display to zero, press the [TARE] button while in RUN mode.
To leave SETUP mode and return to the RUN mode, press [EXIT]. 3.5.3 Moving Through SETUP Menus and Menu Items The following table shows how to move through menus and menu items, and how to display or change the setting of a particular item.
Operating Temp. 5°C to 40°C Relative Humidity Other 80% max. for temperatures up to 30°C; decreasing linearly to 50% max. at 40°C Indoor use at altitudes up to 2000 m; Pollution Degree 2; Overvoltage Category II SPECIAL FEATURES Limits Quantity Digital, isolated control inputs POWER Standard AC powered (automatic selection) Optional Vehicle Powered 5°C to 40°C 5°C to 40°C 5°C to 40°C N/A N/A 4 std., 16 max. 4 4 std., 12 max. 4 4 std., 16 max.
Figure 4-2: Panel Cutout Drawing for DFI 1550 & DFI 1650 (not to scale) • In the panel or rack, cut a hole as shown above. The panel may be up to ¼” in thickness. • Use a 0.062” Allen wrench to remove two setscrews that hold the Panel-Mounting Jacks to the case. To do this, insert the wrench into the side slots at the rear Remove the screws completely. • Remove the Panel-Mounting Jacks by sliding them toward the rear. If the jacks don’t slide easily, tap them gently.
Figure 4-3: Internal Arrangement of DFI 1550 & DFI 1650. Below is a description of each printed circuit board. • The SensoBus Backplane Board serves as the connection between all boards in the instrument. • The Front Panel/Display Board Assembly contains the display and all front panel controls. • The Power Supply Board contains the +15V, -15V, and +5V power supplies. • The Microprocessor Board contains the microprocessor, ROM software, storage chip, and the System connector.
Figure 4-4: External Arrangement of AC powered DFI 1650PT. 4.4.3 Front Panel The pinout for the 25-pin System connector is provided later in this chapter. The pinouts for the individual channels are located in the chapter for that channel. 4.4.4 Case Removal Be sure to remove the power cord from the power source before attempting to remove the instrument from its case. Step1: Remove the 10 Phillips-head machine screws that secure the front panel to the case.
4.5 DFI SC3004 4.5.1 External Arrangement Figure 4-5: External Arrangement of DFI SC3004 4.5.2 Rear Panel The pinout for the 25-pin System connector is provided later in this chapter. The pinouts for the individual channels are located in the chapter for that channel. 4.5.3 Panel Mounting The panel space necessary conforms to the EIA 19” rack-mount standard. Panel mounting ears are attached to the instrument. 4.5.
4.5.7 Internal Arrangement User installable printed circuit boards will slide out of the rear of the case once the case and rear panel have been removed as described above. 4.5.8 Cleaning Turn off the instrument and unplug all connectors. Use a soft cloth or tissue and a mild cleaner. Do not use liquid or aerosol cleaners. Do not allow any cleaner inside the instrument. 4.5.
5.3 Function Input Pins 5.3.1 Overview To use a Function Input pin (9, 10, 11 or 21), connect it to the DGND (pin 19) momentarily. This can be accomplished by a push button switch, relay contact closure, or PLC output. Usually, the Function Input pins perform the default actions described in the “System Connector Pinout”. However, a SensoCode program running on a Mathematics Virtual Channel may replace these default actions. Consult the System Calibration Sheet included with your instrument for details. 5.
Figure 5-2: Open-Collector Output Example 6.0 SYSTEM MENU 6.1 Overview The system menu allows you to examine and change settings that affect the chassis of the DFI instrument. You can view the internal software revision and the instrument’s configuration (i.e. what types of cards are installed in each channel). Detailed instructions on operating the DFI instrument in the SETUP Menu mode can be found in “SETUP Menu mode. A diagram of all menus is located in the “Setup Menu Reference” in Chapter 18. 6.
Installing a channel will cause it to use the “default” or “empty” configuration information for that channel. Any calibration data, SensoCode mathematics programs, display setup, or other information will be erased to default values! All other channels are unaffected. Input or Output Channel Installation Procedure Before installing an Input or Output card, make certain that you know the “card type” (a two-digit hexadecimal number) of the card you wish to install.
12. If the installation was successful, “DONE” will be displayed. If the installation failed, you will see one of the following messages: • “SYSTEM IS FULL”: There are no unused channels available in the instrument. • “WON’T INSTALL”: The Output Relay card type you selected to install does not match the next available set of Limits. For example, the Limit 09-12 card type will not installed unless the Limit 05-08 card type has been installed. • “ERROR 28 ON CH.
7.2 Wiring The System connector on the instrument’s rear panel is used, among other things, for serial communications. See “System Connector”, Chapter 5, for wiring information. All of the serial communications pins on the 25-pin System connector have 500V of electrical isolation from all other pins and connectors on the instrument. Additionally, all serial communications pins are protected against electrostatic discharge (ESD). 7.3 Communications Protocol 7.3.1 RS-232 vs.
7.4.6 TRANSMIT TEST Menu Item When this menu item is selected, the instrument immediately transmits the message “ADDRESS nn TEST” where “nn” is the two-character address of the instrument. Then the message “MESSAGE SENT” is shown on the display. This helps detect wiring or other problems with serial communications from the instrument to the computer or PLC. 7.4.7 LEAVE MENU Item When the menu item is display, press [ENTER] to leave this menu. Press [EXIT] to return to the RUN mode. 8.0 DISPLAY MENU 8.
To change whether the channel’s track, peak or valley data value is displayed on power-up, use the channel’s “OPERATION->POWER->ON SOURCE” menu item. 9.0 LIMITS 9.1 Understanding Limits, Set Points and Return Points Limits are signal levels at which some action (such as a light to come on or go out, or a switch to close) is desired to take place. The point at which this action takes place is the SET POINT.
If no Relay Output channels are installed, there are four Limit menus, which control the operation of the opencollector Limit Outputs on the System connector. Detailed instructions on operating the instrument in the SETUP Menu mode can be found in “SETUP Menu mode”. A diagram of all menus is located in “Setup Menu Reference”. 9.3.2 LIMIT.ENABLE Menu Item This enables or disables the operation of this limit. The two options for this menu item are “ON” and “OFF”. 9.3.3 LIMIT.
Figure 9-2: Limit Operation when LIMIT.ENERGIZE = SIGNAL < SETPOINT Figure 9-3: Limit Operation when LIMIT.ENERGIZE = SIGNAL INSIDE Figure 9-3: Limit Operation when LIMIT.ENERGIZE = SIGNAL OUTSIDE 9.3.6 LIMIT.LATCHING Menu Item This specifies whether to latch the activated limit so that only manually clearing the limit will deactivate it.
9.3.8 LIMIT.SOURCE Menu Item This designates the data source of the channel monitored by this limit. Each channel has three data sources: the live tracking value (TRACK), its highest value (PEAK), and its lowest value (VALLEY). The options for this menu item are: • “TRACK” means the live tracking value of the channel.
7 8 9 10 11 12 (bottom) +OUT -OUT +MEM -MEM AUX1 Analog Output Analog Return (+) Signature (-) Signature/Digital Ground Auxiliary Function 1(connect to pin 10 to activate) 8 10 10 AUX2 Auxiliary Function 2 (connect to pin 10 to activate) 10 The Analog Output and Analog Return pins are electrically isolated from all other pins on the instrument. The maximum recommended cable length to a Signature Module is 100 feet [30m].
10.4 Specifications TRANSDUCER INPUT Transducer type Excitation Voltage Transducer full-scale output Amplifier Gain Selection Calibration Type Differential Input Voltage A/D Converter Low-pass filter Resolution and Frequency Response Full-bridge, strain gage 5 or 10 VDC, short circuit protected with sensing .5 to 21 mV/V @ 5V excitation .5 to 10.5 mV/V @ 10 V excitation Automatic Shunt, mV/V or 2-,3- or 5-point known load +/- 105mV (max.) 24-bit Sigma-Delta Digital, 24-tap FIR See “FREQ.
FREQ.RESPONSE Menu Item This sets the frequency response, step response, and resolution of the channel according to the table below. The tracking value, peak/valley detector values and analog output on this channel are all affected. The default value of 016. HERTZ is suitable for most applications. Frequency Response Step Response (ms) (typical) 002. HERTZ/FAST 002. HERTZ 008. HERTZ 016. HERTZ 032. HERTZ 050. HERTZ 100. HERTZ 250. HERTZ 500. HERTZ 800.
• • • • • “00010” “00020” “00050” “00100” “00200” This menu item may be automatically updated by a transducer’s Signature Module. DISPLAY.UNITS Menu Item Specifies the four-character label that is displayed to the right of the channel’s values. This menu item doesn’t change the mathematical scaling of the channel’s values; that can be changed by altering the “CALIBRATION DATA->FULL SCALE VALUE” menu item and then performing a re-calibration.
Calibration transducers Max. distance of transducer from instrument w/out sense wires Max.
This menu item may be automatically updated by a transducer’s Signature Module. It is available for all calibration types. FULL SCALE VALUE Menu Item The transducer’s full-scale output in engineering units. This menu item may be automatically updated by a transducer’s Signature Module. It is available only with the Shunt Calibration and mV/V Calibration types. ZERO SCALE VALUE Menu Item The transducer’s zero-scale output in engineering units, which is usually zero.
KNOWN POINT x/y Menu Items This enters the engineering units for the known-load calibration points. These points must match the actual loads that you will apply to the instrument during calibration. The number of calibrations points depends on the CALIBRATION TYPE. When using the 2-Point Known Load Calibration type, the following menu items are available: • “KNOWN POINT 1/2”: point 1 of 2, usually 0% of the transducer’s full-scale capacity.
• The display will read WORKING, then APPLY 01000 UNIT (where 01000 and UNIT are previously entered as KNOWN POINT 2/2 and DISPLAY.UNITS). When you have applied this load to the transducer, press [ENTER]. • Upon completion, the display will indicate DONE and the instrument will return to the RUN mode.
If the DAC is set-up to be driven by its own channel’s tracking value, the Analog Output will have the frequency response specified by the channel’s FREQ.ESPONSE menu item. Otherwise, the Analog Output will operate more slowly. DAC.ZERO-SCALE Menu Item This specifies what value, in engineering units, corresponds to zero output on the Analog Output. “Zero output” might be 0 Volts, 2.5. Volts, 5 Volts, 4 mA or 12 mA depending on if the channel has a voltage or current output and how it is configured.
• “YES” will allow a newly detected Signature Module to update the Analog Output settings. • “NO” will now allow automatic updating of the Analog Output settings. 10.5.9 DIAGNOSTICS Sub-Menu DAC FULL SCALE Menu Item When this menu item is selected, the Analog Output of the channel is forced to its zero-scale output, then DAC UPDATED is displayed. This is useful when calibrating or trimming the readout connected to the Analog Output.
10.6.3 Output Selection Jumpers located on the channel’s circuit board determine what outputs are generated when the value selected to drive the Analog Output (from the DAC.CHANNEL and DAC.SOURCE menu items) equals the DAC.FULL SCALE and DAC.ZERO SCALE settings. Figure 10-3: Digital-to-Analog Output Jumper Locations DAC.ZERO SCALE Output DAC.
Analog Output Incorrect Make certain of the type of Analog Output (voltage or current) that the channel is equipped with; see “Identifying the Output Type”. Use the DIAGNOSTICS-DAC FULL SCALE and DIAGNOSTICS-DAC ZERO SCALE menu items to force the Analog Output to a known output. Then, adjust your readout device panel meter, PLC or data acquisition system to match. Check the output selection jumpers; see “Output Selection”.
11.2 Wiring Connect your transducer to an AC/AC-LVDT Input channel by wiring it to the 12-pin connector of that channel. The System Calibration Sheet that shipped with the instrument describes which cards are installed in each channel. The pin-out for this connector is shown on the following table.
Step 4: Determine the LVDT’s electrical null point, then physically mount it. The electrical null is the armature position that produces the least electrical output. Once this point has been determined, the transducer is physically clamped into position. See “Electrical Null and Transducer Mounting” in Section 11.6. Step 5: Perform the calibration. Using the CALIBRATE menu item starts the calibration process. You will be prompted to apply displacements to the transducer as required. 11.
Frequency Response Step Response (ms) (typical) 002. HERTZ/FAST 002. HERTZ 008. HERTZ 016. HERTZ 032. HERTZ 050. HERTZ 100. HERTZ 250. HERTZ 500. HERTZ 800. HERTZ 40 440 110 55 28 16 8 3 2 2 Resolution (counts) (not including minimum 10% overrange/underrange capability) ±25000 ±25000 ±15000 ±10000 ±10000 ±5000 ±5000 ±2000 ±2000 ±2000 Your particular LVDT may have a lower frequency response than the AC/AC LVDT Input Channel.
This menu item doesn’t change the mathematical scaling of the channel’s values; that can be changed by altering the “CALIBRATION DATA-KNOWN POINT X/Y” menu items and then performing a re-calibration. When a character position is flashing press the [UP] or [DOWN] button to change the character. Press [ENTER] to advance to the next character. DISPLAY.AVERAGE Menu Item Controls the speed with which the channel’s display values will update.
KNOWN POINT x/y Menu Items This enters the engineering units for the know-displacement calibration points. These points must match the actual displacements that you will apply to the transducer during calibration. The number of calibrations points depends on the CALIBRATION TYPE. When using the 2-Point Known Displacement Calibration type, the following menu items are available: • “KNOWN POINT 1/2”: point 1 of 2, usually 0% of the transducer’s full-scale capacity.
• The display will read WORKING, then APPLY 01000 UNIT (where 01000 and UNIT are previously entered as KNOWN POINT 3/3 and DISPLAY.UNITS). When you have applied this displacement to the transducer, press [ENTER]. • Upon completion, the display will indicate DONE and the instrument will return to the RUN mode.
11.5.8 DIAGNOSTICS Sub-Menu DAC FULL SCALE Menu Item When this menu item is selected, the Analog Output of the channel is forced to its full-scale output, then DAC UPDATED is displayed. This is useful when calibrating or trimming the readout connected to the Analog Output. DAC ZERO SCALE Menu Item When this menu item is selected, the Analog Output of the channel is forced to its zero-scale output, then DAC UPDATED is displayed.
Step 3: Move the LVDT’s armature to the position that displays the lowest value of raw analog-to-digital converter readings (i.e. a number close to 00.000%). This is the LVDT’s electrical null. Step 4: Firmly clamp down the LVDT’s body. Step 5: Move the LVDT’s armature to roughly the position you wish to define as “positive full scale”.
11.8 Troubleshooting 11.8.1 Error Messages See “Error Messages” in Chapter 17 for information relating to error messages. 11.8.2 Common Problems and Solutions Erratic Display Check electrical connections for continuity and the transducer’s wiring code from its Certificate of Calibration. Make sure that the displacement on the transducer is constant. Check the input to the (+)Signal (“+SIG”) and (-)Signal (“-SIG”) pins with am RMS voltmeter in its “AC Voltage” mode.
The analog-to-digital converter input circuit features adjustable digital, low-pass filtering, 12- to 18-bit resolution (depending on the filter setting) and has several different input ranges. These many ranges allow ±50,000 count resolution (at the slowest filter setting) across a wide variety of input ranges.
Figure 12-1: “Bi-polar Voltage Amp” connection to High-Level Input Channel 12.2.3 “3-wire Voltage” Amplifiers Use the following wiring diagram when connecting an amplified transducer or in-line amplifier with 3-wire voltage amplifier to a High-Level Input channel.
Figure 12-3: “3-wire Voltage Amp w/Single-Wire Shunt Cal” Connection to High-Level Input Channel 12.2.5 “3-wire Current” Amplifiers Use the following wiring diagram when connecting an amplified transducer or in-line amplifier with a 3-wire current amplifier to a High-Level Input channel.
• • • • (+)Excitation supply: “+28 VDC” (-)Excitation supply: “GND” Signal type: “current” Signal reference: “single ended” Figure 12-5: “2-wire Current Amp w/Buffered Shunt Cal” Connection to High-Level Input Channel 12.2.7 “2-wire Current” Amplifiers with Single-wire Shunt Cal Use the following wiring diagram when connecting an amplified transducer or in-line amplifier with a 2-wire current amplifier to a High-Level Input channel.
The High-Level Input channel’s Configuration Jumpers must be set as follows for proper operation. See “Excitation and Signal Jumpers” in Section 12.3. • (+)Excitation supply: “+12 VDC” • (-)Excitation supply: “GND” • Signal type: “voltage” • Signal reference: “single ended” Figure 12-7: “Low Voltage” DC-DC LVDT Connection to High-Level Input Channel 12.3 Excitation and Signal Jumpers 12.3.
Figure 12-7: Excitation and Signal Jumper Locations on the High-Level Input Channel 12.4 Calibration Procedure If you are not familiar with operating the instrument in the SETUP menu mode, see “SETUP Menu mode”. A listing of all menu items is given in “Setup Menu Reference”. Step 1: Wire the transducer to the channel’s connector. Step 2: Set the Excitation and Signal jumpers appropriate for the transducer, amplifier or DC-DC LVDT. Step 3: Enter the CALIBRATION TYPE.
Known-Load Calibration AUXILLIARY INPUTS Quantity Type Response Time Field-Selectable Functions ANALOG OUTPUT Output voltage range Output current range (optional current output channels) Source Isolation Resolution Frequency Response ±0.01% F.S.
500. HERTZ 800. HERTZ 2 2 ±4000 ±2500 ±2500 ±2500 POWER-ON SOURCE Menu Item This menu selects which value is displayed by the channel when first entering the RUN mode. The choices are: • “TRACK” means the live tracking value of the channel. • “PEAK” means the highest value of the channel • “VALLEY” means the lowest value of the channel • “TRACK/PEAK” means the live tracking value and the highest value of the channel.
• “ON” means that the display will update four times each second. The channel’s values will be averaged for ¼ second, then displayed. • “OFF” means that the channel’s display will update as quickly as possible. 12.6.3 AUXn FUNCTION Menu Items The AUX1 FUNCTION and AUX2 FUNCTION menu items determine what happens when the Auxiliary Function pins (labeled as “AUX1” and “AUX2”) on the channel’s connector are activated. These pins are “activated” when connected to the (-)Signature (labeled as “-MEM”) pin.
“KNOWN POINT 4/5” and “KNOWN POINT 5/5” registers. This technique can be used to compensate for the non-linearity in the transducer. The usual loads are zero scale, on-quarter scale, half scale, three-quarters scale and full scale, but you are not restricted to these loads. The selection of the calibration type affects what menu items appear in the CALIBRATION DATA sub-menu. 12.6.
• • • • • “KNOWN POINT 1/5”: point 1 of 5, usually 0% of the transducer’s full-scale capacity. “KNOWN POINT 2/5”: point 2 of 5, usually 25% of the transducer’s full-scale capacity. “KNOWN POINT 3/5”: point 3 of 5, usually 50% of the transducer’s full-scale capacity. “KNOWN POINT 4/5”: point 4 of 5, usually 75% of the transducer’s full-scale capacity. “KNOWN POINT 5/5”: point 5 of 5, usually 100% of the transducer’s full-scale capacity.
• The display will read WORKING, then APPLY 00250 UNIT (where 00250 and UNIT are previously entered as KNOWN POINT 2/5 and DISPLAY.UNITS). When you have applied this load to the transducer, press [ENTER]. • The display will read WORKING, then APPLY 00500 UNIT (where 00500 and UNIT are previously entered as KNOWN POINT 3/5 and DISPLAY.UNITS(. When you have applied this load to the transducer, press [ENTER].
12.6.8 DIAGNOSTICS Sub-Menu DAC FULL SCALE Menu Item When this menu item is selected, the Analog Output of the channel is forced to its full-scale output, then DAC UPDATED is displayed. This is useful when calibrating or trimming the readout connected to the Analog Output. DAC ZERO SCALE Menu Item When this menu item is selected, the Analog Output of the channel is forced to its zero-scale output, then DAC UPDATED is displayed.
12.7.3 Output Selection Jumpers located on the channel’s circuit board determine what outputs are generated when the value selected to drive the Analog Out put (from the DAC.CHANNEL and DAC.SOURCE menu items) equals the DAC.FULL SCALE and DAC.ZERO SCALE settings. Figure 12-8: Digital-to-Analog Output Jumper Locations DAC.ZERO SCALE Output DAC.
Analog Output Incorrect Make certain of the type of Analog Output to a known output. Then, adjust your readout device, panel meter, PLC or data acquisition system to match. Check the output selection jumpers; see “Output Selection”. “APPLY 00000” on Power-up the channel has detected that the transducer connected to the instrument is different than the one the channel was last calibrated with.
In the SYSTEM->CONFIGURATION sub-menu, channels of this type are identified as LIMIT 13-16. 13.2 Wiring The pin-out for the Relay Output channel’s 12-pin connector is shown in the following table.
• “OFF” will turn the relay off. RELAY 3 Menu Item The options for this menu item are: • “ON” will turn the relay on. • “OFF” will turn the relay off. RELAY 4 Menu Item The options for this menu item are: • “ON” will turn the relay on. • “OFF” will turn the relay off. 14.0 DAC OUTPUT CHANNEL 14.1 Features The DAC Output channel uses a digital-to-analog converter to generate a voltage or current from any channel’s track, peak or valley value.
14.4 Specifications Analog Output Voltage Range (voltage output channels) Contact Rating (current output channels) Source Isolation Resolution Update Rate 5, ±5 or 10 VDC 4-20 mA Any channel’s track, peak or valley value 500V 12 bits Depends on how many channels are installed in the instrument 14.5 Channel Menu Each DAC Output channel is configured and calibrated via its channel menu. Detailed instructions on operating the instrument in the SETUP Menu mode can be found in “SETUP Menu Mode”.
14.6 Analog Output Configuration 14.6.1 Identifying the Output Type A DAC Output channel is available with one of two types of digital-to-analog (DAC) outputs: voltage or current. You can determine which type of output a channel has by one of three ways: Consulting the instrument’s System Calibration Sheet. Examining the SYSTEM MENU->CONFIGURATION->CHANNEL nn TYPE menu item where nn is the number of the channel. If the channel’s type is VOLTAGE DAC, it has a voltage output.
Use the DIAGNOSTICS->DAC FULL SCALE and DIAGNOSTICS->DAC ZERO SCALE menu items to force the Analog Output to a known output. Then, adjust your readout device, panel meter, PLC or data acquisition system to match. Check the output selection jumpers; see “Output Selection”. Check the “DAC SETUP->DAC.ZERO-SCALE” and “DAC SETUP->DAC FULL-SCALE” menu items.
• “VALLEY” means the lowest value of the channel since the peak/valley detector was last cleared. RIGHT CHANNEL Menu Item This specifies the channel that is displayed on the right half of the split display. RIGHT SOURCE Menu Item This lets you designate the data source of the channel monitored by the right half of the split display. Each channel has three data sources: the live tracking value (TRACK), its highest value (PEAK), and its lowest value (VALLEY).
16.3 Setup Procedure If you are not familiar with operating the instrument in the SETUP menu mode, see “SETUP Menu Mode”. A listing of all menu items is given in the “Setup Menu Reference” in Chapter 18. Step 1: Examine the System Calibration Sheet. The instrument’s System Calibration Sheet contains important information about what each Mathematics Channel does and what Function Inputs and/or Limit Output pins of the System connector are used. Step 2: Wire to the pins on the System connector.
• “6 DIGIT-UNIPOLAR” displays positive numbers with six full digits (999999) and negative numbers with five full digits (-99999). • “7 DIGIT UNIPOLAR” displays positive number with seven full digits (9999999) and negative numbers with six digits (-999999). DISPLAY.DECPT Menu Item Selects the decimal point location on the channel’s display and serial communications output. Use the [UP] and [DOWN] buttons to move the decimal point to the right and left. DISPLAY.
17.0 ERROR MESSAGES 17.1 Overview If the instrument detects an error during the RUN, INITIALIZE, or SETUP modes, it stops operation and enters its ERROR mode as described in “Operating Modes” in Section 3.1. The most frequent causes of error messages are: • Wiring/electrical connection problems • Misplaced address jumpers when installing or removing channel circuit boards • Swapping channel circuit boards between instruments 17.
Error 37, BAD PERSONALITY Error 38, BAD FORMAT Error 39, WRONG HARDWARE Error 41, SHUNT CAL ERROR MENU-SOFTWARE REV. menu item. The channel doesn’t understand the given card type. Contact Cooper with the information found in the SYSTEM MENU-SOFTWARE REV. menu item. The channel’s EEPROM memory contains a display setting which is not understood by the instrument. Check that every item in the channel’s DISPLAY SETUP menu has a valid setting.
Error 58, WRONG EE VERSION Error 59, CAN’T DO SIG.MOD Error 60, SLAVE TIMEOUT Error 61, LOAD NOT GT LOAD Error 62, SLAVE NAK Error 67, LOADS NOT CLOSE Error 68, CAN’T CALIBRATE Error 70, CAN’T DO GAIN Error 71, SLAVE CHECKSUM Error 72, CAN’T DO FILTER Error 73, SLAVE NOT INIT Error 74, SLAVE BAD REPLY Error 75, SLAVE NOT FOUND Error 81, OPERAND EXPECTED Error 82, BAD OPERAND Error 83, BAD INSTRUCTION Error 84, BAD VALUE Error 85,UNREACHABLE GOTO Error 87, GOTO NOT JUMP CF 66 below.
18.0 SETUP MENU REFERENCE 18.1 Navigation instructions • • • • • Press [UP]+[DOWN] to enter the SETUP menu mode. Press [UP] to move up. Press [DOWN] to move down. Press [ENTER] to select an item. Press [EXIT] to restart the instrument.
Table 18-2: System Menu CF 66 73 Rev.
Table 18-3: LIMIT MENU Table 18-4: SERIAL COMMUNICATIONS MENU Table 18-5: DISPLAY MENU for models DFI 1650, DFI 1650 PT Table 18-6: DISPLAY MENU for DFI SC3004 CF 66 74 Rev.
Table 18-7: STRAIN GAGE INPUT CHANNEL MENU CF 66 75 Rev.
Table 18-7: STRAIN GAGE INPUT CHANNEL MENU (Continued) CF 66 76 Rev.
Table 18-8: AC-AC LVDT INPUT CHANNEL MENU CF 66 77 Rev.
Table 18-8: AC-AC LVDT INPUT CHANNEL MENU (Continued) CF 66 78 Rev.
Table 18-9: HIGH-LEVEL INPUT CHANNEL MENU CF 66 79 Rev.
Table 18-9: HIGH-LEVEL INPUT CHANNEL MENU (Continued) CF 66 80 Rev.
Table 18-10: RELAY OUTPUT CHANNEL Table 18-11: DAC OUTPUT CHANNEL MENU Table 18-12: SPLIT DISPLAY VIRTUAL CHANNEL CF 66 81 Rev.
Table 18-13: MATHEMATICS VIRTUAL CHANNEL CF 66 82 Rev.
WARRANTY REPAIR POLICY LIMITED WARRANTY ON PRODUCTS Any of our products which, normal operating conditions, proves defective in material or in workmanship within one year from the date of shipment by Cooper, will be repaired or replaced free of charge provided that you obtain a return material authorization from Cooper and send the defective product, transportation charges prepaid with notice of the defect, and establish that the product has been properly installed, maintained, and operated within the limi