Specifications
9
High-Performance Autosampler
High-Sensitivity Application Data (Analysis of Selenium (Se))
Mix and inject up to four sample types (diluted solution, standard solution,
sample, matrix modifier, etc.). (Of course, unmixed samples can also be injected.)
Select a fluororesin tube or pipette tip as the injection nozzle.
Automatic dilution and re-analysis if the calibration curve concentration is exceeded.
Independent control of the gas flow through the graphite tube during
atomization achieves extremely high-sensitivity measurements.
(Japanese Patent 2067563)
e
cted.)
e
xceeded.
Digital temperature control and electronic gas flow control enhance stability
Combination of a high-sensitivity optical sensor and unique digital temperature control
technology achieves highly accurate temperature control in all temperature regions from
drying to atomization. An electronic flow controller can precisely control the inner gas flow rate
in 0.01L/min units. These control techniques significantly enhance both the sensitivity and the
data stability.
H
i
g
h-Performance Autosampl
er
H
i
g
h-Sensitivity Application Data (Analysis of Selenium (Se)
)
Example of Automatic Dilution and Re-Analysis
Analysis of cadmium (Cd): The maximum concentration of the calibration curve has been set to 1ppb. At 1.8ppb, the sample of [003] has exceeded the
maximum concentration. As a resultit has been automatically diluted by 4 times and re-measured.
* Depending on the state of the sample, some other combinations may be appropriate.
High-density graphite tube
(P/N 206-50587)
Pyro-coated graphite tube
(P/N 206-50588)
Platform tube
(P/N 206-50887-02)
Selecting the Graphite Tube
Effective for elements that readily form
carbides (Ni, Fe, Cu, Ca, Ti, Si, V, Mo, etc.).
Restricts chemical interference due to
coexisting substances. Effective for the analysis
of environmental samples and biosamples,
such as sea water and industrial waste.
Can be used for all elements. Especially
effective for low-boiling point elements
(Cd, Pb, Na, K, Zn, Mg, etc.).
Measurement Example Showing High Sensitivity
Analysis of 0 to 5ppb selenium (Se): Sensitivity is adequate for 1
ppb measurements.
(20μL injected volume, Pb modifier, pyro-coated graphite tube)
Measurement Example Showing High Stability
Analysis of 0.5ppb lead (Pb): Approx. 5% relative standard deviation
after 5 repeated measurements is adequate for quantitative analysis.