Specifications

25
* Notes
.
The PC, monitor and printer are not included in the standard configuration. Prepare them separately.
.
The ASC-7000, ASK-7000 and GFA-7000 are optional accessories.
.
The separately available micro sampling kit is required for the flame micro sampling method.
.
Microsoft Windows Vista Business or XP Professional is required to run the AA-7000 Series software.
.
Microsoft Windows Vista Business and XP Professional are registered trademarks of Microsoft Corp., U.S.A.
.
The AA-7000 Series does not conform to IEC60601 –– Medical Electrical Equipment.
Flame Continuous Method
Required Software Operation Environment
Furnace Method / Flame Micro Sampling Method
Autosampler
Common Specifications
Functions Zero-point detection, auto diagnosis, auto rinse, random access
Sample volume Sample vials: 15 mL, reagent vials: 53 mL
Solvent rinse aspiration method
Sampler: 340 W × 280 D × 270 H mm, 8 kg
Controller (including stand): 260 W × 200 D × 320 H mm, 7 kg
Nozzle rinse
Dimensions and weight
Functions
Zero-point detection, auto diagnosis, auto rinse, auto mixing, random access
Sample volume Sample vials: 2 mL, reagent vials: 20 mL
Solvent rinse discharge method
Sampler (including sliding parts): 340 W × 280 D × 400 H mm, 11 kg
Controller: 235 W × 110 D × 240 H mm (built into GFA-7000), 6 kg
Nozzle rinse
Dilution function, reagent addition function
Sampling functions
250 μL
Syringe
2 to 90 μL
Injection volume
1%RSD (20 μL)
Repeatability
Rinse port: Less than 0.0001
Mixing port: Less than 0.00001
Carryover
Solvent rinse discharge method, solvent discharge and rinse with next sample
Mixing port rinse
Performed in mixing port. Maximum mixture volume: 0.6 mL
Mixing function
Up to 4 solutions
Number of solutions possible for mixing
.
Calibration curve method: 5 solutions max. (sample + 4 types of reagents)
.
Standard addition method: 6 solutions max. (sample + standard solution + 4 types of reagents)
Maximum number of
mixing reagents added
For measurement result on unknown samples:
.
If extrapolation of calibration curve is possible: automatic calculation of dilution rate and dilution to bring concentration within calibration curve range
.
If extrapolation of calibration curve is not possible: dilution rate fixed at 10x
Auto dilution / re-analysis
Dimensions and weight
Intel
®
Celeron 420 (1.60 GHz or higher)
To use GFA-TV, Intel
®
Pentium DualCore E2180 (2 GHz) or higher is required.
1 GB or higher (Vista) or 512 MB or higher (XP)
To use GFA-TV, 1 GB or higher (GFA-TV)
XGA (1024 × 768 dots) or higher
One CD-ROM drive (for installing a software)
Requires 60 MB min. hard disk space for AA installation.
To use GFA-TV, a separate partition of HDD (HDD inside PC) from than that for OS and AA measurement data storage is recommended to
use for movie recording.
One serial port (for AA control)
USB 2.0 port (for GFA-TV)
CPU
Microsoft Windows Vista Business / XP Professional
Operating system
RAM
Monitor
Storage device
I/O port
Monitor, keyboard, mouse, printerPeripheral device
Shimadzu recommends the PC and monitor in the PC Set for AA-7000 Series with WizAArd Software Pre-Installed.
Purchase a printer separately. The personal computer and monitor specifications are as follows.
RS-232C communication control from AA unitControl
Maximum reagent / sample positions
Rinse water bottle
Power requirements
Reagents: 8 positions Samples: 60 positions (random access available with reagents or samples)
2 L
Choose from 100, 120, 220, or 230 VAC, 50 VA, 50/60 Hz