Technical Information
Technical Information:
Accuracy: The accuracy of the
C
ONTOUR blood glucose monitoring
system has been assessed in a correlation study conducted by health-
care professionals. In this study, 108 fresh capillary blood specimens
were tested with the
C
ONTOUR blood glucose monitoring system by
108 untrained people with diabetes using three test strip lots. The
reference measurement was done on the YSI Glucose Analyzer.
C
ONTOUR data compare well with the results from the YSI method.
Regression Statistics (Plasma/Serum Reference):
plasma reference glucose range: 75 to 397 mg/dL
Lot y =
Correlation
Coeffi cient (r)
1 1.02x – 0.12 mg/dL 0.972
2 1.03x – 4.34 mg/dL 0.963
3 0.95x – 0.41 mg/dL 0.974
Precision: A study was conducted with the
C
ONTOUR blood glucose
monitoring system with five heparinized whole blood specimens with
glucose levels from 40 to 325 mg/dL. Multiple replicates (n=100)
were tested using multiple
C
ONTOUR blood glucose meters and one
lot of
C
ONTOUR blood glucose test strips. The following precision
results were obtained:
Plasma/Serum:
Mean SD
Coeffi cient
of variation
38 mg/dL 1.9 mg/dL 4.8%
101 mg/dL 5.1 mg/dL 5.1%
118 mg/dL 4.4 mg/dL 3.7%
205 mg/dL 6.8 mg/dL 3.3%
326 mg/dL 11.7 mg/dL 3.6%
Your CONTOUR blood glucose meter is specific for glucose and has
been referenced to plasma/serum values. Methods calibrated to
plasma/serum can be 9 to 15% higher than the results of whole blood
glucose assay methods.