For proper disinfection, you must keep all endcap surfaces
wet for 60 seconds.
®
Buttons
USB Connector (under cap)
Strip Port
Disinfecting Your Meter
Supplies Needed for Disinfecting:
• Germicidal wipes containing 0.55% sodium hypochlorite
(bleach)
• Paper towels
• Timing device
1. Before disinfecting, clean the meter as described in
Cleaning Your Meter.
For proper disinfection, you must keep all meter surfaces
wet for 60 seconds.
Using a new germicidal wipe, carefully wipe all outer surfaces
of your meter until wet. Do not allow cleaning solution to run
into the meter through areas such as around the buttons or the
meter’s test strip or data ports.
2. Dry all meter surfaces and test strip port using a clean paper
towel if needed.
Disinfecting Your Lancing Device
Supplies Needed for Disinfecting:
• Germicidal wipes containing 0.55% sodium hypochlorite
(bleach)
• Paper towels
• Timing device
1. Before disinfecting, clean the lancing device as described
above in Cleaning Your Lancing Device.
Cleaning Your Lancing Device
Supplies Needed for Cleaning:
• Germicidal wipes containing 0.55% sodium hypochlorite
(bleach)
• Paper towels
1. Clean the M
ICROLET
®
2 lancing device with germicidal wipes
until all soil is removed.
2. Dry as necessary with a clean paper towel.
3. Using a new germicidal
wipe, carefully wipe all
external and internal
surfaces of the endcap until
wet.
For proper disinfection, you must keep all lancing device
surfaces wet for 60 seconds.
2. Remove the endcap and
lancet, if present.
Cleaning and Disinfection
Cleaning and Disinfection
84766780_CntrNextLINK_UG_ENca_FpBp_v0.indd 68-6984766780_CntrNextLINK_UG_ENca_FpBp_v0.indd 68-69 3/27/13 7:59 AM3/27/13 7:59 AM