User guide

TECHNICAL
AND CARE
64
TECHNICAL
AND CARE
65
Symbols Used
The following symbols are used throughout the product labeling
for the CONTOUR
®
NEXT LINK blood glucose monitoring system
(meter packaging and labeling, and reagent and control solution
packaging and labeling).
System Specifications
Test Sample: Capillary whole blood
Test Result: Referenced to plasma/serum glucose
Sample Volume: 0.6 μL
Measuring Range: 1.1 mmol/L to 33.3 mmol/L
Countdown Time: 5 seconds
Memory: Stores most recent 1,000 test results
Battery Type: Non-serviceable, 250mAh rechargeable
lithium polymer battery, 3.4V – 4.2V (5V input voltage)
Meter/Battery Life: 5 years
Charging Current: 300 mA
Strip Storage Temperature Range:
C
30ºC
Normal Control Storage Temperature Range:
C
30ºC
Meter Operating Temperature Range:
C
45ºC
Humidity: 10%–93% RH
Dimensions: 97 mm wide x 31 mm high x 18 mm thick
Weight: 43 grams
Sound Output: 45 to 80 dBA at a distance of 10 cm
Electromagnetic Compatibility (EMC): The
CONTOUR
®
NEXT LINK meter complies with the
electromagnetic requirements specified in ISO 15197: 2003.
Electromagnetic emissions are low and unlikely to interfere
with other nearby electronic equipment, nor are emissions
from nearby electronic equipment likely to interfere with
the CONTOUR NEXT LINK meter. Immunity to electrostatic
discharge meets the requirements of IEC 61000-4-2: 2008.
It is advisable to avoid use of electronic devices in very dry
environments especially if synthetic materials are present.
The CONTOUR NEXT LINK meter has been tested for radio
frequency interference at the frequency range and test levels
specified by ISO 15197: 2003. To avoid radio frequency
interference, do not use the CONTOUR NEXTLINKmeter near
cellular or cordless telephones, walkie talkies, garage door
openers, radio transmitters or other electrical or electronic
equipment that are sources of electromagnetic radiation, as
these may interfere with the proper operation of the meter.
®
Symbols Used
System Specifications
Symbol
Symbol
What It Means What It Means
Use by date
(last day of month)
Recycle Packaging
Batch code Control Range Normal
C
45ºC
Temperature
limitations
Control Range Low
Control Range High
Consult instructions
for use
Do not reuse
In Vitro Diagnostic
Medical Device
Sterilized using
irradiation
Catalogue number Caution
Batteries must be disposed of in accordance with laws
in your country. Contact your competent local authority
for information on the relevant laws regarding disposal
and recycling in your area.
The meter should be treated as contaminated and
disposed of according to local safety rules. It should
not be disposed of with waste electronic equipment.
The USB cable is waste electrical and electronic
equipment. Do not dispose in household waste.
Remember to respect local regulations.
Contact your health care professional or local waste disposal
authority for medical waste disposal guidelines.