User guide

USING THE
MAINMENU
40
USING THE
MAINMENU
41
Enter this Meter IDEnter this Meter ID
into your pump:into your pump:
123456123456
?
NextNext
Go BackGo Back
CONNECT TO PUMPCONNECT TO PUMP
METER ID MENU
Add ID
Delete ID
Review ID
REVIEW METER ID
1. 123456
2. ------
3. ------
15. After the meter and pump are connected, you set your
Send Options
. This controls how your meter sends blood
glucose readings to your Medtronic pump. Choose the
Send Option
that is best for you.
Your options are:
Always
Meter will always send results
Ask Me
Meter will ask you before sending results
Never
Meter will never send results
Automatically send
results to the Pump?
Always
Ask me
Never
SEND OPTIONS
8. Scroll to
Connect Devices
and press
ACT
.
9. Select
On
and press
ACT
. If you are
replacing
a meter, go to
Step 10
. If
you are
adding
a new meter, go to
Step 12
.
NOTE: For detailed instructions, see your compatible
Medtronic device user guide.
Then scroll to
Meters
and press
ACT
.
NOTE: If your UTILITIES MENU does not contain Connect
Devices, scroll to Meter Options and press ACT.
UTILITIES MENU
Alarm Clock
Connect Devices
Block
DEVICE OPTIONS
Meters
Remotes
Other Devices
METER OPTION
Off
On
METER ID MENU
Add ID
Delete ID
Review ID
DELETE METER ID
1. aaaaaa
2. bbbbbb
3. ------
METER ID MENU
Add ID
Delete ID
Review
ADD METER ID
123456
12. To add a new meter, select
Add ID
and
press
ACT
. Refer to Step 6 (or see
meter back) to find meter ID (SN).
Use the up and down arrow buttons to
enter the six character meter ID shown
on the meter screen.
Press
ACT
after each character entry.
®
Connecting the Meter and Pump
Connecting the Meter and Pump
13. Select
Review ID
on your compatible
Medtronicdevice
to confirm that your
CONTOUR
®
NEXTLINK
meter ID has been
correctly entered. If
incorrect, go back to
Step 10.
14. Return to your
C
ONTOUR NEXT LINK
meter. After you have
entered the meter ID
into your compatible
Medtronic device,
press
Next
.
To change your choice,
see page 42.
10. To replace a meter, select
Delete ID
and
press
ACT
.
11. If you have multiple meter IDs, repeat
Step 10 for each one.