Operation Manual
Medical Test Supply www.medicaltestsupply.com
23
Chapter21 NIBP Specification
Name Electronic Sphygmomanometer
Display 2.8'' color LCD Display
NIBP Specifications
Measurement Method Oscillometric method
Working modes Automatic
Measurement Range
Pressure
adult
0~290mmHg(0~38.6kPa)
pediatric
0~235mmHg(0~31.3kPa)
neonatal
0~140mmHg(0~18.6kPa)
Pulse: 40~240/min
Inflation
adult 160mmHg
pediatric 120mmHg
neonatal 70mmHg
Alarm Range
adult mode
SYS ALM: 40~270 mmHg
DIA ALM: 10~215 mmHg
pediatric mode
SYS ALM: 40~200 mmHg
DIA ALM: 10~150 mmHg
neonatal mode
SYS ALM: 40~135 mmHg
DIA ALM: 10~100 mmHg
Overpressure protect
adult mode 295±5mmHg
pediatric mode
240±5mmHg
neonatal mode
145±5mmHg
Resolution
Pressure 1mmHg
Measurement Accuracy
Cuff Pressure Accuracy ±3mmHg
Error
The BP Value of the device is equivalence with the measurement
value of Stethoscope. The error meets all the conditions in the
ANSI/AAMI SP-10:2002+A1:2003 +A2:2006.
Operating
Temperature/ Humidity
+5ºC~40 ºC . 15%RH~80%RH
Transport and Storage
Temperature/Humidity
-20ºC~+55ºC . ≤95%RH
Atmospheric pressure 80kPa~105kPa
Battery 4 "AA" alkaline batteries, AC Adapter separately sold
Main Unit Dimensions 130(L)*110(W)*80mm(H)
Main Unit Weight 300gram