Chapter21   NIBP Specification 
Name  Electronic Sphygmomanometer 
Display  2.8'' color LCD Display 
NIBP Specifications 
Measurement Method  Oscillometric method 
Working modes  Automatic 
Measurement Range 
Pressure 
adult 
0~290mmHg(0~38.6kPa) 
pediatric 
0~235mmHg(0~31.3kPa) 
neonatal 
0~140mmHg(0~18.6kPa) 
Pulse: 40~240/min 
Inflation 
adult  160mmHg 
pediatric  120mmHg 
neonatal  70mmHg 
Alarm Range 
adult mode 
SYS ALM: 40~270 mmHg 
DIA ALM: 10~215 mmHg 
pediatric mode 
SYS ALM: 40~200 mmHg 
DIA ALM: 10~150 mmHg 
neonatal mode 
SYS ALM: 40~135 mmHg 
DIA ALM: 10~100 mmHg 
Overpressure protect 
adult mode  295±5mmHg 
pediatric mode 
240±5mmHg 
neonatal mode 
145±5mmHg 
Resolution
Pressure  1mmHg 
Measurement Accuracy 
Cuff Pressure Accuracy  ±3mmHg 
Error 
The BP Value of the device is equivalence with the measurement 
value of Stethoscope. The  error meets all  the  conditions in  the 
ANSI/AAMI SP-10:2002+A1:2003 +A2:2006. 
Operating 
Temperature/ Humidity 
 +5ºC~40 ºC   . 15%RH~80%RH 
Transport  and  Storage 
Temperature/Humidity 
 -20ºC~+55ºC   . ≤95%RH 
Atmospheric pressure  80kPa~105kPa 
Battery  4 "AA" alkaline batteries, AC Adapter separately sold 
Main Unit Dimensions  130(L)*110(W)*80mm(H)  
Main Unit Weight  300gram 