Chapter20 Specification
Name Electronic Sphygmomanometer
The degree of
protection against
ingress of liquid
Ordinary sealing device, without the function of preventing liquid from
entering
Display mode 2.8"color LCD Display
NIBP Specifications
Measurement Method Oscillometric method
Working mode Manual operation
Measurement Range
Pressure
adult 0~290mmHg(0~38.6kPa)
pediatric 0~235mmHg(0~31.3kPa)
neonatal 0~140mmHg(0~18.6kPa)
Pulse: 40~240/min
Inflation
adult 160mmHg
pediatric 120mmHg
neonatal 70mmHg
Overpressure protect
adult mode 295±5mmHg
pediatric mode 240±5mmHg
neonatal mode 145±5mmHg
Resolution Pressure: 1mmHg(0.133kPa)
Accuracy Static pressure: ±3mmHg(±0.4kPa)
Error
The BP Value of the device is equivalence with the measurement value
of Stethoscopy. The error meets all the conditions in the ANSI/AAMI
SP-10:2002+A1:2003 +A2:2006.
Operating Temperature/
Humidity
+5ºC40 ºC 15%RH80%RH
Transport
Transport by general vehicle or according to the order contract, avoid
pounded, shake and splash by rain and snow in transportation.
Storage
Temperature: -20ºC+55ºC; Relative humidity: 95%; No corrosive
gas and drafty.
Atmospheric pressure 700hPa1060hPa
Battery 4 "AA" alkaline batteries, AC Adapter(AC, 100V-240V, optional)
Dimensions 130(L)*110(W)*80mm(H)
Unit Weight 300g(not including batteries)