Chapter16 Maintenance, Cleaning and Keeping
*Please follow the instructions in the user manual. If you do not comply, our company will
not assume responsibilities of the quality.
Warning
Must remove batteries and cut off the alternating power supply before cleaning the
sphygmomanometer, accessories must be clean separately.
Caution
Do not high pressure disinfect the sphygmomanometer and accessories.
Do not immerse the sphygmomanometer and accessories into any liquid.
Do not use any sphygmomanometer and accessories that may be damaged or deteriorated.
Cleaning:
If there is something dirty on the sphygmomanometer and accessories, please wipe it with a
dry and soft cloth.
If the sphygmomanometer and accessories are very dirty, you can clean them with a soft
cloth dampened with water or neutral detergent after the cloth can be full twisted.
Maintenance:
Advice
Do not use any naphtha, thinner or gas to clean.
Do not attempt to clean or wash the cuff.
Frequently clean the Sphygmomanometer and accessories.
The device should be calibrated regularly( or according to
the calibration procedure of hospital). Users can go to the
inspection agency designated by the state to calibrate it, or
compared with the standard pressure gauge, at any measuring
point, the maximal error of measuring the pressure in cuff
should be ±0.4kPa(±3mmHg); also can contact with our
company.