User`s manual

66
• Switch OFF the power and wait till the Control Console is turned off. Remove the Dead Man Switch
Connector Jig which is connected at the bottom of the Base Unit.
• Connect the exposure switch cable to the J3 connector on the power board.
• Place the Power Switch to the ON position.
• Check functioning by performing trial exposures.
• With successful completion of trial exposures, the equipment is ready to use.
9.3 Exposure history through Console:
• Switch off the power to the unit.
• Switch ON the unit and press "UP" & "DOWN" keys simultaneously within 2
seconds after the logo appears then the screen changes to "CONFIGURA-
TION MENU" as shown in Figure 130 . Press "DOWN/Decrement" key till
"Service" option gets highlighted.
• Using the Control Console press the keys Molar followed by bitewing,
canine, premolar, molar, canine, bitewing and S/mA as shown in Figure
129 .
• Then display changes to "SERVICE" screen as shown in Figure 134 . Press
"Down/Decrement" key till "Exposure count" gets highlighted.
• Note the total exposures shown in the report and Switch off the unit.
Figure 134
9.4 Preventive Maintenance
• Recommended once in a year (after 1st year of usage) :-
• Check all the Scissor Arm movements & adjust as per “5.14 Mechanical Adjustments”.
• Inspect all the cables & replace wherever defective.
• Inspect all the fasteners for tightness & tighten accordingly.
• In case of non-usage for long period (>6months) :-
• X-Ray Tube Seasoning has to be carried out.
9.5 Disposal of the Unit
• Some parts of the equipment contain material and fluids which must be disposed off in special areas
designated by the local health authorities or other local regulations at the end of the equipment’s life
cycle.
• The Manufacturer and the Distributor do not accept any responsibility for the disposal of equipment or
parts discarded by the user and the related costs.
• All parts that need to be disposed shall be listed in the Service report with a comment "The following
parts are handed over to <customer name> for disposal as per local health authorities or other local
regulations" during handing over to Customer.
• In particular the equipment contains the following materials and / or components:
• Tube head: external packages in non-biodegradable plastic, dielectric oil, lead, copper, brass, alumi-
num, tungsten.
• Power supply and remote control: external packages in non biodegradable plastic, iron, populated
printed circuit boards, copper.
• Tube head extension: Iron, Aluminum, Copper & Silicon rubber.