User`s manual
30
5 Installation of intraskan dc
5.10 Scissor-Arm Operation Checking
Move the Scissor arm and ensure there is no drift as per steps given below.
• Keep the Vertical Arm in vertical position and horizontal arm in horizontal position.
• Move the horizontal arm down (folding movement) in small jerk free incremental steps of approximate-
ly 10 degrees.
• At each step as above, leave the scissor arm and ensure that there is no drift movement. Continue till
the horizontal arm reaches vertical position.
• Keep the Vertical Arm vertical and horizontal arm horizontal position.
• Pull the L Arm in small jerk free incremental steps of approximately 10° so that the horizontal arm stays
horizontal always and the Vertical arm moves down.
• At each step as above, leave the scissor arm and ensure that there is no drift movement. Continue till
the maximum expanded position of the scissor arm is reached.
• Keep the Scissor arm in folded position.
• Refer “5.14 Mechanical Adjustments” in case of minor adjustments of scissor-arm.
5.11 Other Checks:
• Ensure that the Straight-Arm rotation is 170-190° approximately from end to end of the Base Unit.
• Ensure that the Tube head rotation about the axis is 300-310° approximately.
• Ensure that the L arm rotation angle is 520-540° approximately.
• Ensure that the scissor-arm rotation about the straight-arm is 200-210° approximately.
5.12 Power On Check:
• For new Installation, if installation date is > 6 months of manufacturing date of Tube Head, carry out
Tube seasoning procedure (5.13 X-Ray Tube Seasoning) before proceeding.
• Switch On the power and give the following exposures using internal console and exposure switch.
Also give exposures using external switch configuration if installed.
• 70kV/6mA, 2 Sec Exposure 2 Times
• 55kV/4mA 40 ms Exposure 2 Times
• If no errors are reported, then the Unit is ready to use.
5.13 X-Ray Tube Seasoning
• In case of non-usage for long period (>6months) X Ray Tube Seasoning is recommended. Cover the
Cone with Lead Cap. Using the Control Console set the parameters as per Table below. Give Exposure
and repeat exposure 5 times for each combination of kV, mA and ms. After all the exposures are com-
pleted, the Unit is ready for use.
Table 5.14-1: Tube Seasoning Protocol
Kv mA Time (ms) Kv mA Time (ms) Kv mA Time (ms)
55 4 40 60 4 40 70 4 40
55 6 40 60 6 40 70 6 40
55 8 40 60 8 40 70 8 40
55 4 500 60 4 500 70 4 500
55 6 500 60 6 500 70 6 500
55 8 200 60 8 200 70 8 200