User`s manual
21
5 Installation of Intraskan dc
5.6 Ground Connection Check:
Using a Multimeter, check the continuity between the following points:-
• Between Ground point of power board and Tube Head inner Cone metal part,
• Between Ground point of power board and Scissor Arm Guide rod and
• Between Ground point of power board and Base Unit Wall mounting plate as shown in Figure 28 .
• If any of the check fails then check the Ground connections inside Base Unit & Tube Head for Cable fault.
Figure 28: Check the continuity
of ground connection.
1
2
3
5.7 Remote console Installation:
• Take the remote console template from packing box and fix it at the desired location on the stud.
• Confirm the level of the template using spirit level. Mark the locations for the 2 mounting holes on the
template and mark the wiring hole also at the locations shown in Figure 29 .
• Drill the 2 mounting holes marked using 5/32" drill bit to the depth of 1.26" and drill the direct wiring
hole(>1.5" dia).
Figure 29: 2
Mounting holes
Remote Console Template