Specifications
Data Sheet
Horizontal System Analog Cha nn els
Characteristic MSO5034
DPO5034
MSO5054
DPO5054
MSO5104
DPO5104
MSO5204
DPO5204
Maximum Sample Rate (All
channels)
5GS/s 5GS/s 5GS/s 5GS/s
MaximumSampleRate(1or2
channels)
——
10 GS/s 10 GS/s
Maximum Equivalent Time
Sampling Rate
400 GS/s
Maximum Record Length with
Standard Configuration
12.5M
12.5M (4 ch)
25M (1 or 2 ch)
Maximum Record Length with
Option 2RL
25M
25M (4 ch)
50M (1 or 2 ch)
Maximum Record Length with
Option 5RL
50M
50M (4 ch)
125M (1 or 2 ch)
Maximum Record Length with
Option 10RL
125M
125M (4 ch)
250M (1 or 2 ch)
Maximum Duration at Highest
Real-Time Sample Rate
25 ms
Time Base Range
250
ps/div to 1000 s/div
Tim
e Resolution (in ET/IT mode)
2.5
ps/div
Time Base Delay Time Range –5 divisions to 5000 s
Ch
annel-to-Channel Deskew
Ra
nge
±75 ns
TriggerJitter(RMS) 100 fs
RMS
with enhanced triggering ON
100 ps
RMS
with enhanced triggering OFF
Time Base Accuracy ±5 ppm over any 1 ms interval
Horizontal System Digital Channels
Characteristic All MSO5000 Models
Maximum Sample Rate
(Main)
500 MS/s (2 ns resolution)
Maximum Record
Length (Main)
12.5M Standard
Up to 40M with Record Length options
Maximum Sample Rate
(MagniVu)
16.5 GS/s (60.6 ps resolution)
Maximum Record
Length (MagniVu)
10k points centered around the trigger
Minimum Detectable
Pulse Width
1ns
Channel-to-Channel
Skew (typical)
200 ps
Maximum Input Toggle
Rate
500 MHz at minimum input swing; higher toggle rates
can be achieved at higher amplitudes
Trigger System
Characteristic
Description
Main Trigger Modes
Auto, Normal, and Single
Trigger Coupling DC, AC, HF Rej (attenuates >50 kHz), LF Rej
(attenuates <50 kHz), Noise Reject (reduces sensitivity)
Trigger Holdoff Range
250nsto8s
Trigger Sensitivity
Internal DC Coupled For 1 M:
1mV/divto4.98mV/div:0.75divfromDCto50MHz,
increasing to 1.3 div at instrument bandwidth
5 mV/div: 0.40 div from DC to 50 MHz, increasing to
1 div at instrument bandwidth
For 50 :
0.40divfromDCto50MHz,increasingto1divat
instrument bandwidth
External (Auxiliary Input)
1M
200 mV from DC to 50 MHz, increasing t o 500 mV at
250 MHz
Trigger Level Rang e
Any Channel ±8 divisions from center of screen
External (Auxiliary Input)
±8 V
Line
Fixed at about 50% of line voltage
14 www.tektronix.com