Technical data
August 2006 CS–4001
Manual
Page 7
5. Removing the connections from CS/TMC-4001
First you must disconnect all crocodile-clamps from the battery.
Next you can remove the connections C1 to C8 from CS/TMC-4001.
You must not switch off the CS/TMC-4001 before all connections are removed
from the battery.
6. Technical Data
No. of measuring inputs
CS-4001/127 TMC-4001/63
Cell/Block voltage 127 63
Total voltage 2 1
Current 2 1
Temperature 2 none
Accuracy
Input Range DC Resolution Accuracy Input impedance
3V 1,00 mV
±0,05% ±2 Digits
Cell/Block voltage
15V 1,00 mV
±0,05% ±2 Digits
75V 10,0 mV
±0,10% ±2 Digits
Total voltage
600V 100 mV
±0,10% ±2 Digits
≥ 1 MΩ
60mV 0,01 mV
±0,10% ±2 Digits
Current (shunt-voltage)
1000mV 0,10 mV
±0,10% ±2 Digits
≥ 900 kΩ
600V CAT II
Measuring multi-cables
If not otherwise specified (when ordering), CS/TMC-4001 instruments are equipped with standard
measuring multi-cables (Type MKB16/250) specified for 250V. Measuring multi-cables specified for
500 V (Type MKB16/500) are also available on request.
Others
Mains supply 85-264V~, 47-63Hz
Operating temperature +10°C – 35°C
Storage temperature +5°C – 50°C