Specifications
Table Of Contents
- Imprint
- Documentation Representative
- © HYDAC FILTER SYSTEMS GMBH
- Contents
- Preface
- Safety information
- Storing the CS
- Decoding the model code label
- Checking the scope of delivery
- CS1000 Features
- CS1000 Restrictions on use
- CS1x1x dimensions (without display)
- CS1x2x dimensions (with display)
- Hydraulic connection types
- Fastening / mounting the CS1000
- Display rotatable/Adjustable As Needed
- CS1000 hydraulic installation
- Electrical connection of the CS1000
- Setting the measuring mode
- Operating the CS1x2x using the keypad
- Overview of menu structure
- Using switching output
- Setting limit values
- Reading the analog output
- Status Messages
- Connecting CSI-D-5 (Condition Sensor Interface)
- Connecting the CS1000 to an RS-485 bus
- Communicating with the CS1000 via the RS-485 bus
- Taking the CS1000 out of operation
- Disposing of CS1000
- Spare Parts and Accessories
- Cleanliness classes - brief overview
- Checking/resetting default settings
- Technical data
- Recalibration
- Customer Service
- Model Code
- EC declaration of conformity

ContaminationSensor CS 1000 Reading the analog output
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 84/112
BeWa CS1000 3764916 300 en-us 2012-08-29.doc 2012-08-29
Fluid temperature TEMP
The current range 4.8 … 19.2 mA or voltage range 2.4 … 9.6 V is dependent on the
fluid temperature of -25°C … 100°C (resolution: 1°C) or -13°F … 212°F (resolution:
1°F) as shown in the table below.
Current I = Temperature / Error Voltage U =
I < 4,00 mA Cable break U < 2,00 V
4,0 mA < I < 4,1 mA Device error, device not ready 2,00 V < U < 2,05 V
4,1 mA < I < 4,3 mA Not defined 2,05 V < U < 2,15 V
4,3 mA < I < 4,5 mA Flow error
(The flow rate is too low.)
2,15 V < U < 2,25 V
4,5 mA < I < 4,8 mA Not defined 2,25 V < U < 2,40 V
I = 4,8 mA -25 °C / -13 °F U = 2,40 V
… … …
I = 7,68 mA 0 °C / 32 °F U = 3,84 V
I = 8,26 mA +5 °C / 41 °F U = 4,13 V
I = 8,83 mA +10 °C / 50 °F U = 4,42 V
I = 9,41 mA +15 °C / 59 °F U = 4,70 V
I = 9,98 mA +20 °C / 68 °F U = 4,99 V
I = 10,56 mA +25 °C / 77 °F U = 5,28 V
I = 11,14 mA +30 °C / 86 °F U = 5,57 V
I = 11,71 mA +35 °C / 95 °F U = 5,86 V
I = 12,29 mA +40 °C / 104 °F U = 6,14 V
I = 12,86 mA +45 °C / 113 °F U = 6,43 V
I = 13,44 mA +50 °C / 122 °F U = 6,72 V
I = 14,02 mA +55 °C / 131 °F U = 7,01 V
I = 14,59 mA +60 °C / 140 °F U = 7,30 V
I = 15,17 mA +65 °C / 149 °F U = 7,58 V
I = 15,74 mA +70 °C / 158 °F U = 7,87 V
I = 16,32 mA +75 °C / 167 °F U = 8,16 V
I = 16,90 mA +80 °C / 176 °F U = 8,45 V
I = 17,47 mA +85 °C / 185 °F U = 8,74 V
I = 18,05 mA +90 °C / 194 °F U = 9,02 V
I = 18,62 mA +95 °C / 203 °F U = 9,31 V
I = 19,20 mA +100 °C / 212 °F U = 9,60 V
19,2 mA < I < 19,8mA Not defined 9,60 V < U < 9,90 V
19,8 mA < I < 20 mA No measured value 9,90 V < U < 10 V