Specifications
Table Of Contents
- Imprint
- Documentation Representative
- © HYDAC FILTER SYSTEMS GMBH
- Contents
- Preface
- Safety information
- Storing the CS
- Decoding the model code label
- Checking the scope of delivery
- CS1000 Features
- CS1000 Restrictions on use
- CS1x1x dimensions (without display)
- CS1x2x dimensions (with display)
- Hydraulic connection types
- Fastening / mounting the CS1000
- Display rotatable/Adjustable As Needed
- CS1000 hydraulic installation
- Electrical connection of the CS1000
- Setting the measuring mode
- Operating the CS1x2x using the keypad
- Overview of menu structure
- Using switching output
- Setting limit values
- Reading the analog output
- Status Messages
- Connecting CSI-D-5 (Condition Sensor Interface)
- Connecting the CS1000 to an RS-485 bus
- Communicating with the CS1000 via the RS-485 bus
- Taking the CS1000 out of operation
- Disposing of CS1000
- Spare Parts and Accessories
- Cleanliness classes - brief overview
- Checking/resetting default settings
- Technical data
- Recalibration
- Customer Service
- Model Code
- EC declaration of conformity

ContaminationSensor CS 1000 Using switching output
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 52/112
BeWa CS1000 3764916 300 en-us 2012-08-29.doc 2012-08-29
Using switching output
You can use the switching output in the modes described below. For a further
description of the measurement modes, see page 28.
Mode M1: Continuous measurement
Purpose: Measurement only
Function: Continuous measurement of cleanliness class Switching function
only for "Device ready".
Mode M2: Continuous measurement and switching
Purpose: Continuous measurement and controlling of signal lamps etc.
Function: Continuous measurement of solid contamination, continuous
monitoring of programmed limit values; the switching output is
enabled and switches on the monitoring display or alarm on site
Mode M3: Filter to cleanliness class and stop
Purpose: Clean up hydraulic reservoir
Function: Control of a filter unit, continuous measurement of solid
contamination. If pre-programmed cleanliness level is reached 5
times in sequence, the pump is stopped.
Mode M4: Filter to continuously monitor cleanliness class
Purpose:
Establish continuous monitoring of cleanliness class between
min/max limit values
Function: If min/max limit values are pre-programmed, the CS switches the
filter unit on/off to keep cleanliness within the limit value range
Load the switching output with a maximum of 2 A and 30 V DC.
Mode "SINGLE" measurement
Purpose: Perform a single measurement and "stop" the result.
Function: Single measurement of solid contamination without switching
functions Switching function only for "Device ready".