Specifications
Table Of Contents
- Imprint
- Documentation Representative
- © HYDAC FILTER SYSTEMS GMBH
- Contents
- Preface
- Safety information
- Storing the CS
- Decoding the model code label
- Checking the scope of delivery
- CS1000 Features
- CS1000 Restrictions on use
- CS1x1x dimensions (without display)
- CS1x2x dimensions (with display)
- Hydraulic connection types
- Fastening / mounting the CS1000
- Display rotatable/Adjustable As Needed
- CS1000 hydraulic installation
- Electrical connection of the CS1000
- Setting the measuring mode
- Operating the CS1x2x using the keypad
- Overview of menu structure
- Using switching output
- Setting limit values
- Reading the analog output
- Status Messages
- Connecting CSI-D-5 (Condition Sensor Interface)
- Connecting the CS1000 to an RS-485 bus
- Communicating with the CS1000 via the RS-485 bus
- Taking the CS1000 out of operation
- Disposing of CS1000
- Spare Parts and Accessories
- Cleanliness classes - brief overview
- Checking/resetting default settings
- Technical data
- Recalibration
- Customer Service
- Model Code
- EC declaration of conformity

ContaminationSensor CS 1000 Setting the measuring mode
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 28/112
BeWa CS1000 3764916 300 en-us 2012-08-29.doc 2012-08-29
Setting the measuring mode
Once the sensor is switched on or supplied with power, it automatically runs in the
measuring mode that has been set.
Mode M1: Continuous measurement
Application: Stand-alone sensor
Data output: Display & RS485 & analog output
Purpose: Measurement only
Function: Continuous measurement of cleanliness class Switching
function only for "Device ready".
Mode M2: Continuous measurement and switching
Application: Stand-alone sensor with alarm standby display
Data output: Display & RS485 & analog output & switching output
Purpose: Continuous measurement and controlling of signal lamps etc.
Function: Continuous measurement of solid contamination, continuous
monitoring of programmable limit values; switching output is
activated to switch on the monitor display or an alarm on site
Mode M3: Filter to cleanliness class and stop
Application: Controlling a filter unit
Data output: Display & RS485 & analog output & switching output
Purpose: For cleaning up a hydraulic reservoir
Function: Control of a filter unit, continuous measurement of solid
contamination. If pre-programmed cleanliness level is achieved
5 times in sequence, the pump is stopped.
Load the switching output with a maximum of 2 A and 30 V DC.