Parameters
All parameters are set to read and write by any level of user unless otherwise stated.
HIGH TEMPERATURE ALARM SET POINT (not supported in Field Trial Prototype units)
Description
Parameter
Range
Increment &
Units
Default
The Absolute Temperature above which an alarm is triggered.
The measured temperature must exceed this setpoint for a
time specified by Temperature Out-Of-Spec Alarm Delay
before the alarm is triggered.
-40.0 to 15.0
or disabled
0.1°C
disabled
LOW TEMPERATURE ALARM SET POINT
(not supported in Field Trial Prototype units)
Description
Parameter
Range
Increment &
Units
Default
The Absolute Temperature below which an alarm is triggered.
The measured temperature must exceed this setpoint for a
time specified by Temperature Out-Of-Spec Alarm Delay
before the alarm is triggered.
-40.0 to 15.0
or disabled
0.1°C
disabled
TEMPERATURE OUT-OF-SPEC ALARM DELAY (not supported in Field Trial Prototype
units)
Description
Parameter
Range
Increment &
Units
Default
The amount of time the measured temperature must be
continuously above or below the alarm setpoints before an out-
of-spec alarm is triggered.
If the measured temperature returns to values between the
high and low alarm setpoint range, this time counter is reset.
0.5 to 4
hours
0.5 hours
2 hours
DOOR OPEN DELAY (not supported in Field Trial Prototype units)
Description
Parameter
Range
Increment &
Units
Default
The time between the cabinet door opening (and staying
open), and the Door Open Alarm triggering.
0 seconds to
1800
seconds
10 seconds
180
seconds
(3
minutes)