Owner's manual
Configuration and Calibration 23
Figure 5-9. Set-Up Mode Selection
5.13 I/O Set-Up Mode Selection
Method 1
I/O (SELECT) - See Section 5.4 on page 13.
Method 2
I/O (DIRECT) - Closing only rocker switch position
3 on switch S1 only places the indicator directly into the I/O
mode. See Section 5.3 on page 12.
The I/O SET-UP mode accesses the configuration of available inputs and outputs in the DLR. There are five main
groups of configurable inputs and outputs that are defined as follows:
• Serial I/O - configures ports 1 and 2 for baud rate, parity, continuous or demand, etc.
•Time and date - sets the onboard clock for time and date (allows time and date to be transmitted in the
serial output).
•Parallel output - configures DLR3110 for BCD output. This is an option requiring the installation of the
parallel output option board.
•Analog output - configures DLR3110 for analog output (4-20 mA or 0-10 VDC). This is an option
requiring the installation of the analog output option board.
•Discrete input and outputs - configures two discretes to function as either control output(s) or remote
command input(s).
Press
UP or DOWN arrow keys to step through selections. Press LEFT or RIGHT arrow keys to change the current
state of the selection. The modes may be exited at any time by pressing the
CE key. If the CE key is pressed, the
prompt
Exit? YES or NO scrolls continuously until the YES or NO key is pressed.
I/O ?
ZERO
DISPLAY
TARE PRINT
YES
NO
CE ENTER