5.6.1 Calibration Mode Menu
Press LEFT or RIGHT arrow keys to change values. The various menus may be exited or data selection aborted at
any time by pressing the
CE key. If the CE key is pressed, the prompt Exit ? YES or NO scrolls continuously until
either the
YES or NO key is pressed.
*XX represents existing data.
**XXX represents the present capacity of the DLR3110.
Use the following steps to set calibration data in the DLR3110 using the Calibration mode menu.
1. Verify that the INPUT valve (1) is closed (rotate clockwise until it stops) and that the VENT valve (2) is
open (two turns counter-clockwise from its stop). See Figure 5-1 on page 12.
2. Press the
ENTER key on the display panel keypad. The LED display reads 0.00. Close the VENT valve
(rotate clockwise until it stops).
NOTE: If ZERO is not accepted, display shows either SIGNAL INPUT UNDER MINIMUM ZERO RANGE or
INPUT ABOVE CAL INPUT RANGE (NO CAL messages).
3. Set the input pressure from the test standard to the target value on the display. Wait for system to stabilize
and press the ENTER key on the display panel.
NOTE: For pneumatic setup, see Section 5.1 on page 12. For absolute models, use a vacuum pump to reach
the applicable SPAN target values that are below local barometric pressure.
LED Display Alpha-Numeric Display Key Pressed Remarks
**XXX CAPACITY DOWN arrow To change the capacity value, see
Section 5.7 on page 19.
**XXX X .5 To change the resolution value, press LEFT
or RIGHT arrow key.
BASE PSI Press LEFT or RIGHT arrow key to select
the base unit used to calibrate indicator. If
for any reason this selection is changed, the
indicator must be calibrated to the new
base unit selection.
---- GAGE ?
or ABS ?
Press LEFT or RIGHT arrow key to select
choice.
*XX ZERO CAL See step 1 below Does not exist on ABS type units. Disregard
steps 1 and 2 below on ABS units.
*XX SPAN 1 TARGET VALUE=XX See step 3 below To change the target value, see Section 5.7
on page 19.
Table 5-6. Calibration Mode Parameter Selections
CAL ?
ZERO
DISPLAY
TARE PRINT
YES
NO
CE ENTER