16 DLR3110 Rack Mount Digital Pressure Calibrator Manual
5.5.3 Auto Set-Up Menu (Absolute)
See Section 9.0 on page 63 for a complete listing of the default settings for all gage and absolute selections under
the AUTO set-up mode.
NOTE: If you continue to press NO key past highest full scale range, DLR3110 will go to exit menu. Hit YES key at exit
menu (see Table 5-3 on page 14).
5.6 Calibration Set-Up Configuration Mode Selection
Use one of the following methods for configuring the calibration set-up mode:
Method 1
CAL (SELECT). See Section 5.4 on page 13 for information on selecting main mode settings.
Method 2
CAL (DIRECT). Closing only rocker switch position
2 on switch S1 places the DLR3110 directly into
CALIBRATION mode. See Section 5.3 on page 12.
NOTE: If the AUTO SET-UP selection is accepted and completed, the indicator automatically advances to the calibration
mode (see Section 5.5 on page 14).
LED Display Alpha-Numeric Display Key Pressed Remarks
SET AUTO ? YES key See Section 5.5 on page 14
---- GAGE ? NO key See Section 5.5 on page 14
— ABS ? YES key See Section 5.5 on page 14
SET-UP 15 PSIA NO key Continue to press NO key until the
alpha-numberic display shows the
DLR3110’s full scale range.
YES key Press YES key when the alpha-numeric
display shows the DLR3110’s full scale
range.
SET-UP ACCEPT ?
YES or NO
NO key Continue to press NO key until the
alpha-numberic display shows the
DLR3110’s full scale range. See note
below.
——YES key Accept data
XXXX CAPACITY YES key “XXX” shown on LED display represents
the full scale range and is based on
previous selection.
— EXIT ? YES key Return to SET-UP/ I/O?. (see Section 5.4
on page 13).
NO key Return to XXX/CAPACITY
Table 5-5. Auto Set-Up Mode (Absolute) Parameters Selections