Owner's manual

14 DLR3110 Rack Mount Digital Pressure Calibrator Manual
Figure 5-3. Auto Set-Up Mode
5.5 Auto Set-Up Configuration Mode Selection
The auto set-up automatically sets the following conditions in the DLR3110:
SET-UP - CONFIG configuration parameters
SET-UP - CONV conversion parameters
SET-UP - I/O configuration of inputs/outputs
---- CAL sets capacity, count by, abs/gage, and cal units; resets all calibration values; sets all cal target
values for span and hysteresis.
5.5.1 Auto Set-Up Menu (General)
* Press YES key to accept selection; NO key to reject.
Upon completion and accepting an AUTO set-up configuration, the program automatically advances the
indicator to the calibration mode (see Section 5.6.1 on page 17).
LED Display Alpha-Numeric Display Key Pressed* Remarks
SET AUTO ? YES key See Section 5.4 on page 13
---- GAGE ? YES key See Section 5.5.2 on page 15
NO key Continue to ---/ABS?
---- ABS ? YES key See Section 5.5.3 on page 16
NO key Continue to ---/SPECIAL?
---- SPECIAL. ? NO key Allocated for future use, continue to
----/CUSTOM ?
---- CUSTOM ? NO key Allocated for future use, continue to
----/EXIT ?
---- EXIT ? YES key Return to main set-up mode selections
(see Section 5.4 on page 13).
NO key Returns to ----/GAGE ?
Table 5-3. Auto Set-Up Mode Selection Parameters
AUTO ?
ZERO
DISPLAY
TARE
PRINT
YES
NO
CE
ENTER