Datasheet

Page 6 of 28
Spectrum Master™ MS2712E and MS2713E Specifications
Channel Scanner (Option 0027)
Number of Channels 1 to 20 Channels
Measurements Graph/Table, Max Hold (On/5 sec/Off), Freq/Channel, Current/Max, Single/Dual Color
Scanner Scan Channels, Scan Frequencies, Scan Customer List, Scan Script Master™
Amplitude Reference Level, Scale
Custom Scan Signal Standard, Channel, # of Channels, Channel Step Size, Custom Scan
Frequency Range 100 kHz to 4 GHz (MS2712E),
100 kHz to 6 GHz (MS2713E)
Frequency Accuracy ± 10 Hz + Time base error
Measurement Range –110 dBm to +26 dBm
Application Options Bias-Tee (On/Off), Impedance (50 Ω, 75 Ω, Other)
CW Signal Generator Option (Option 0028) (Requires Option 0021)
(Requires CW Signal Generator Kit, P/N 69793)
Setup Parameters
Frequency Frequency, Signal Standard, Channel Number, Display Setup Help
Amplitude Power Level (Low/High), Offset (dB)
Frequency Range 2 MHz to 2 GHz
Frequency Reference Accuracy: ± 1.5 ppm (25 °C ± 25 °C) + aging, < ± 50 ppb with GPS On
Output Power High 0 dBm typical, Low –30 dBm typical
Attenuator (included in kit 69793): 0 to 90 dB in 1 dB steps
Gated Sweep (Option 0090)
Mode Spectrum Analyzer, Sweep
Trigger External TTL
Setup Gated Sweep (On/Off)
Gate Polarity (Rising, Falling)
Gate Delay (0 ms to 65 ms typical)
Gate Length (1 µs to 65 ms typical)
Zero Span Time
20 MHz BW Demod (Option 0009)
Required for all signal analyzers except AM/FM/PM Signal Analyzer, Option 509