Datasheet
Spectrum Master™ MS2712E and MS2713E Specifications
Spectrum Analyzer (continued)
Spectral Purity
SSB Phase Noise @ 1 GHz -100 dBc/Hz, -110 dBc/Hz typical @ 10 kHz offset
-105 dBc/Hz, -112 dBc/Hz typical @ 100 kHz offset
-115 dBc/Hz, -121 dBc/Hz typical @ 1 MHz offset
Amplitude Ranges
Dynamic Range > 102 dB (2.4 GHz), 2/3 (TOI-DANL) in 1 Hz RBW
Measurement Range DANL to +26 dBm
Display Range 1 dB to 15 dB/div in 1 dB steps, ten divisions displayed
Reference Level Range -120 dBm to +30 dBm
Attenuator Range 0 dB to 55 dB in 5 dB steps
Maximum Continuous Input +30 dBm
Amplitude Units Log Scale Modes: dBm, dBV, dBmv, dBμV
Linear Scale Modes: nV, μV, mV, V, kV, nW, μW, mW, W, kW
Amplitude Accuracy
100 kHz to 4.0 GHz ± 1.25 dB, ± 0.5 dB typical
> 4.0 GHz to 6 GHz ± 1.50 dB, ± 0.5 dB typical
Displayed Average Noise Level (DANL)
Preamp Off Preamp On
(Reference level -20 dBm) (Reference level -50 dBm)
(RBW = 1 Hz, 0 dB attenuation) Maximum Typical Maximum Typical
10 MHz to 2.4 GHz -141 dBm -146 dBm -157 dBm -162 dBm
>2.4 GHz to 4 GHz -137 dBm -141 dBm -154 dBm -159 dBm
>4 GHz to 5 GHz -134 dBm -138 dBm -150 dBm -155 dBm
> 5 GHz to 6 GHz -126 dBm -131 dBm -143 dBm -150 dBm
Spurs
Residual Spurious < -90 dBm (RF input terminated, 0 dB input attenuation, > 10 MHz)
Input-Related Spurious < -75 dBc (0 dB attenuation, -30 dBm input, span < 1.7 GHz, carrier offset > 4.5 MHz)
Exceptions, typical < -70 dBc @ <2.5 GHz, with 2072.5 MHz Input
< -68 dBc @ F1 - 280 MHz with F1 Input
< -70 dBc @ F1 + 190.5 MHz with F1 Input
< -52 dBc @ 7349 - (2F2) MHz, with F2 Input, where F2 < 2424.5 MHz
< -55 dBc @ 190.5 ± (F1/2) MHz, F1 < 1 GHz
Third-Order Intercept (TOI)
Preamp Off (-20 dBm tones 100 kHz apart, 10 dB attenuation)
800 MHz +16 dBm
2400 MHz +20 dBm
200-2200 MHz +25 dBm, typical
> 2.2 GHz to 5.0 GHz +28 dBm, typical
> 5.0 GHz to 6.0 GHz +33 dBm, typical
Second Harmonic Distortion
Preamp Off, 0 dB input attenuation, -30 dBm input
50 MHz -56 dBc
> 50 MHz to 200 MHz -60 dBc, typical
> 200 MHz to 3000 MHz -70 dBc, typical
VSWR
2:1, typical
Page 3 of 28