User's Manual

Table Of Contents
25
BTTracer Protocol Analyzer Users ManualCATC Version 1.0
Analyzer Setup screen.
You can select only one item at this point. If both the BusEngine and the
Firmware need to be updated, the update will complete for the first item
and then return to the above screen so the second update can be
performed.
When the second update has finished, you will see the following message
telling you that the update is complete.
Step 5 Power cycle BTTracer to complete the update.
Manual Updates to BusEngine and Firmware
You can manually update BTTracers Firmware and/or BusEngine by
performing the following steps:
Step 1 Select Analyzer under Setup on the Menu Bar.
You see the Analyzer Setup screen:
Step 2 Click Reset Analyzer
The Analyzer resets, performs self-diagnostics, and returns to service.
Note The self-diagnostics should complete about five seconds after the trigger LED
lights. If the diagnostics fail, the trigger LED blinks on and off continually,
indicating faulty hardware. If this occurs, contact CATC for customer support.