User's Manual

Table Of Contents
Procedure
1.
Activate a job.
2. From the Transient main menu, press ALT > F5 Data Display Settings > F8 Show Tach.
7.6 Set the number and type of plots to display
Select up to four plots to display transient data. If you want to select fewer than four plots,
select an option below and choose the Plot Off option. The option you select is used for all
plots in Transient.
Procedure
1. Activate a job.
2. From the Transient main menu, press ALT > F6 Plot Setup.
3. For each plot type listed, use the up and down arrows to select the type of plot to
display.
The options vary based on the number of channels you set. The analyzer can display
an orbit from any two inputs (for example, AB, AC, AD, BA). For four channels, the
analyzer uses AB and CD for the orbits.
4. Press Enter.
5. Repeat until all options are set as desired.
7.7 Set data collection parameters
7.7.1 Setup/Acquire Data
The Setup/Acquire Data option on the Transient main menu lets you set the data collection
parameters for your job, including:
The Fmax
The sample rate, number, and time
PeakVue technology or Demodulation
The tachometer
Trigger values for a tachometer
Data collection actions for overload conditions
Sensors
You can also set display parameters from the Transient ALT2 main menu.
Advanced Transient
155
DRAFT