User's Manual

Table Of Contents
High Resolution Analysis
Bearing/Gear Analysis- PeakVue
Low Frequency Analysis - SST
Coast Down Peak Hold
Order Tracking
Synchronous Analysis
You can also listen to vibration without collecting any data. The F3 Listen To Live Data on the
Analyze main menu lets you listen without collecting data. When you select a
measurement, the audio turns off.
Note
The audio is not stored or recorded.
Prerequisites
Pair a Bluetooth device. See Section 2.11.2.
You must have the Analysis Experts help enabled to collect and listen to vibration
data. See Section 6.8.2.
Procedure
1.
From the Analyze main menu, do one of the following:
Press F3 Listen To Live Data.
Select an Analysis Expert, and press F6 Listen To Live Data.
Display a waveform or spectrum plot and press F6 Listen To Live Data.
The Bluetooth Listener screen appears.
2. Set up the following options as necessary.
Option Description
F1 Volts Accel Set the audio signal to use the Accel or the Volts input. The default is the
input for the active measurement point when you selected Listen To Live
Data.
F2 Input Set the audio signal to use an input channel (A, B, C, D). The default is the
input for the active measurement point when you selected Listen To Live
Data.
F6 Test Headset Test the Bluetooth communication between the analyzer and the
headphones. Use your headphones to listen to the sound that is generated
for several seconds.
F8 Filter Set the filter for the audio signal. You can select Off (no filter), 2000 Hz High
Pass, or 8000 Hz High Pass.
F9 Gain Increase the amplitude of the audio signal. The signal gain affects the
volume. You can also use the up arrow key.
Analyze and Advanced Analyze
139
DRAFT