2-32 Basic Analyzer Operations
Backlight – This field is only available in the UltraSpec 8117 analyzer. It is
used to turn the analyzer’s backlight on and off. When on, the backlight
turns off automatically after one minute of key inactivity and back on auto-
matically when any key is pressed.
Load Firmware
This feature is used to load firmware and applications from the PC into the
analyzer. See “Installing the Application” on page 1-10 for complete
instructions on this procedure.
Review Calibration
To ensure maximum measurement accuracy, each UltraSpec analyzer is
thoroughly calibrated by CSI before shipping, and the results of the cali-
bration are recorded on the calibration tag that is shipped with the analyzer.
The values entered into the analyzer’s calibration routine allow the firm-
ware to compensate for any minor errors in measurement that may be
caused by component tolerances within the analyzer.
Normally, these values do not have to be modified. However, the analyzer
calibration values should be checked against the calibration tag under the
following conditions:
• The analyzer produces invalid or unreasonable readings.
• The battery has been allowed to discharge to the point where ana-
lyzer memory has been lost. (This typically would require several
months of inactivity.)
• New firmware has been loaded into the analyzer (UltraSpec 8000
analyzer only).
• The analyzer displays the “Warning – Check Calibration Factors”
screen.
Normally, the calibration values can simply be entered into the analyzer
using the values recorded on the calibration tag. It is recommended, how-
ever, that the analyzer be returned to CSI once a year for re-calibration in
order to maintain the instrument’s accuracy.