User Manual

1-29
Communicating with UltrMgr
Note
The option in UltraMgr to “Download Jobs to Analyzer” and
“Download a Default Job” is not available for the UltraSpec
Analyze and UltraSpec Motor applications. The “Download
Tolerance Table” option is available only in the UltraSpec
Alignment applications.
Note
There is no communication between UltraMgr and the
UltraSpec Thermal Growth application.
Interrupting the Job Loading Process
If an interruption should occur during the job load process, the analyzer
will attempt to reestablish communications with the computer for a prede-
termined length of time. This time allotment, designated “Timeout,” is con-
figurable in the Communications Setup menu accessed from the
Communications Menu (see “Communications Setup” on page 2-22). CSI
recommends that this be set to 10. After timeout, the analyzer will display
a Communications Error screen. Press the Reset/Main key, check the fol-
lowing items, and repeat the job load procedure.
If problems or error messages are encountered during the communication
process, recheck the following:
1.····The baud rate and stop bit setting on both the analyzer and the
computer.
2. ···The cable connection between the analyzer and the computer.
3. ···The date and time on both the analyzer and the host computer.
4. ···The communications port setup on the computer.
5. ···The number of retries and time out settings on both the analyzer and
the computer.
6. ···The communications cable being used is the proper communications
cable furnished with the analyzer.