Manual Chapter 5

6-36 Acquiring Alignment Data
Note
In this example, Quick Spec uses the Auto Sweep method to
acquire data.
22
The next screen directs you to turn on the laser heads, position them at any
angular position you desire to start from, and then press the Enter key to
initialize the laser heads. It is not required to define a sweep arc range and
direction of rotation due to the increased number of data readings acquired
every 2 degrees and the process, which automatically defines the direction
of rotation. Once the laser heads are initialized the next screen appears
prompting you to now acquire the data by rotating the laser heads. After
the laser heads have been rotated, press the Enter key transmit or transfer
the data from the laser heads to the analyzer.
Note
For the Auto Sweep method, if the laser heads have not been
turned off since they last acquired and stored a set of data read-
ings and prior to initializing them, pressing the Alt/0 (Alt/
Space) key will “Resend” the last set of data they acquired and
stored to the analyzer without having to resweep the laser
heads. This is useful if for some reason the data transfer from
the heads to the analyzer was aborted before it was complete.