When the data transfer is completed, an “Analysis In Progress” screen sim-
ilar to the following is displayed while the data is curve fit to a sine wave. If
the fit is satisfactory (85% and above), you are returned to the main menu
where the data may now be viewed.
18
19
Otherwise, you are warned the data is “Unfit” (less than 85%). At this point
you should either repeat the data acquisition or try to manually condition
or edit the data. This built-in check helps alert you to the data losing reli-
ability in the misalignment calculations. Refer to “Additional Data Detail”
on page 5-34 for more information.