Manual Chapter 5
5-17
Job Definition
Laser Configuration
It does not matter which sensor head is put on which machine, however,
the analyzer must know each head’s location. Use any keypad key to toggle
from one configuration to the other. The arrows must be set to match the
directions that each laser will transmit, i.e., the arrows must represent the
actual laser directions by pointing away from the small circle and toward
the red target. Press Enter to accept the configuration.
Caution!
Selecting the proper configuration for the laser heads is extremely
important! If the setting is wrong, all of the machine move calculations
will be incorrect.
For the Model 8215/8225, A and B are marked on the face plate of the laser
head.
Note
The program automatically knows whether Model 8215 or
8225 laser heads are being used.