Manual Chapter 1
2-22 Basic Analyzer Operations
Note
The transfer is performed via direct computer to analyzer
hook-up using a Model 639 communications cable. This cable
is supplied as standard equipment with the analyzer (no
modem is required). When Host PC Load/Dump is selected,
the analyzer automatically awaits signals from the computer to
begin transferring stored job(s) data. A screen stating that the
analyzer is awaiting host commands will be displayed. This
screen will be displayed until the time needed for all retries has
been used up. At that point, a communications error message
will be displayed.
Communications Setup
The Communications Setup menu is used to configure the communications
port of the analyzer.
7
Select the Baud Rate from 76.8K, 38.4K, 19.2K, 9600, 4800, 2400, 1200, or
300 baud. For rapid data transfer, select the highest baud rate possible, con-
sistent with the capabilities of the host computer equipment.
Set the Stop Bit option at One or Two.
The Timeout option controls the amount of time (selectable from 0 through
99.9 seconds) that the analyzer will wait for the computer to respond before
aborting the procedure and trying again. Entering a value of zero will dis-
able this feature.