Manual Chapter 1

1-24 Read This First
Note
Select this option whether you are replacing the base firmware
or
loading an application or program.
The following screen will be displayed.
14
9.··· Select Yes and press the Enter key to continue.
Loading Firmware
Caution!
Reloading firmware in an UltraSpec 8000 analyzer may reset calibra-
tion factors. Always review calibration factors after loading firmware
to verify the settings are the same as in the factory supplied calibration
card. Refer to “Review Calibration” for information about calibrating
the UltraSpec analyzer.
There may be up to a 20-30 seconds pause before loading begins. After the
firmware has been loaded, a list of application programs to be loaded will
be shown (if available on the same diskette). An error message will appear
if no application programs are located.
10. · If the error message occurred, insert the Application Diskette into the
computer’s drive A (or B). On the analyzer, press the Reset/Main
key. Otherwise, go directly to Step 11.