Manual Chapter 4
4-33
Attaching the Fixtures
Starting the UltraSpec Analyzer
Turn on the UltraSpec analyzer by pressing the ON/OFF button located at
the top right of the Front Panel. This automatically takes you to the pro-
gram’s main menu. From the main menu, press the Options key. Select (4)
Alignment Setup and press Enter.
From the Alignment Setup menu, configure the alignment parameters.
(Refer to “Alignment Setup” on page 3-6 for details.) The default settings
are:
Method
– Auto Sweep
ACQ Mode
– Standard
Tolerance
– Angle/Offset
Thermal Growth
– No