User's Guide

C h a p t e r 6 S p e c i f i c a t i o n s
  

Guidance and manufacture’s declaration electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should assure that it is used in such an environment.
Immunity
test
IEC
60601
test
level
Compliance
level
Electromagnetic environment - guidance
Conducted
RF
IEC
61000-4-6
3 V
rms
150 kHz
to 80
MHz
6 V
rms
in ISM
band
3 V
rms
6 V
rms
in ISM band
Portable and mobile RF communications
equipment should be used no closer to any part
of the device, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
𝒅 = 𝟏, 𝟐
𝑷
Radiated
RF
IEC
61000-4-3
3 V/m
80 MHz
to 2.7
GHz
3 V/m
𝒅 = 𝟏, 𝟐
𝑷 𝟖𝟎 𝑴𝑯𝒛 𝒕𝒐 𝟖𝟎𝟎 𝑴𝑯𝒛
𝒅 = 𝟐, 𝟑
𝑷 𝟖𝟎𝟎 𝑴𝑯𝒛 𝒕𝒐 𝟐, 𝟕 𝑮𝑯𝒛
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol: