C h a p t e r 6 S p e c i f i c a t i o n s
Guidance and manufacture’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of COMPUMEDICS should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floor is covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for a.c.
Power Ports
±2 kV for d.c.
Power Ports
±1 kV for
input/output
lines
±2kV for a.c.
Power Ports
±2 kV for d.c.
Power Ports
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
±1 kV differential
mode
Mains power quality should
be that of a typical
commercial or hospital
environment.