S e r v i c e a n d M a i n t e n a n c e C h a p t e r 4
4.2 Periodic Maintenance and Inspection
No regular maintenance or calibration is required for this
device, although a regular annual inspection is recommended.
Grounding continuity, leakage current and isolation should be
checked periodically by a qualified technician. Where
appropriate such tests should conform to any standard
applicable to the country in installation.
Inspection should also consist of checking than any plugs are
engaged and not damaged. Any cords should not be damaged
or knotted. Damage to the enclosure or any parts of the device
should be evaluated by the manufacturer.
Frequency of inspection and cleaning is dependent on the
location type and use and should be determined by a qualified
engineer, the equipment maintenance department or
personnel. Inspect the device at a minimum of once per year.
Technical information such as diagrams and parts list are
available on request from service@compumedics.com