15.POWER REQUIREMENTS
15.1 List of fuse specifications
Fuse values to be installed on-site [A]
220 – 240 V110 – 120 V
LightL
F1
[A]
Battery
F2/F3
[A]
L
F1
[A]
Battery
F2/F3
[A]
D 500plus
D 530plus
210410
D 650plus4166,316
D 500
D 530
210410
D 6504166,316
All fuse slow blow
15.2 List of terminal voltages
The terminal voltages specified in table below must be provided on the flanged tube of the CHROMOPHARE®
light or the Power Box
Ceiling tubePower Box
LightA.C.
[V]
D.C.
[V]
A.C.
[V]
D.C.
[V]
D 500plus
D 530plus
21,5 (*24 + 1,5/-0,321,5 (*24 + 1,5/-0,3
D 650plus21,5 (*24 + 1,5/-0,321,5 (*24 + 1,5/-0,3
D 500
D 530
21,5 (*24 + 1,5/-0,321,5 (*24 + 1,5/-0,3
D 65021,5 (*24 + 1,5/-0,321,5 (*24 + 1,5/-0,3
(* filtering board next in line