Service manual

61
CHROMOPHARE® D 650plus, D 530plus, D 500plus, D 650, D 530, D 500 Service Manual (E)
9.TEST AND ADJUSTMENT POINTS
9.1 Test points CHROMOPHARE® D 650
TP 1 Terminal block 24 V D.C. loaded, tolerance + 1.5 / - 0.3 V
TP 2 Sliding contact 24 V D.C. loaded, tolerance + 1.5 / - 0.3 V
TP 3 Sliding ring 23.8 V D.C. loaded, tolerance + 1.5 / - 0.3 V
TP 4 Sliding ring 23.6 V D.C. unloaded, tolerance + 1.5 / - 0.3 V
TP 5 Cable connectors input X1/2 to X1/4 22.6 V A.C. / D.C., tolerance + 1.5 / - 0.3 V
TP 6 Main bulb between cable 8 and 6 max. 22.0 V A.C. / D.C. RMS, tolerance + 0.3 V
Setting of the max. bulb voltage see page 63 - 64, by means of adjustable resistor R 20
9.2 Test points CHROMOPHARE® D 650plus, D 530plus, D 500plus, D 530, D 500
TP 1 Terminal block 24 V D.C. loaded, tolerance + 1.5 / - 0.3 V
TP 2 Sliding contact 24 V D.C. loaded, tolerance + 1.5 / - 0.3 V
TP 3 Sliding ring 23.8 V D.C. loaded, tolerance + 1.5 / - 0.3 V
TP 4 Sliding ring 23.6 V D.C. unloaded, tolerance + 1.5 / - 0.3 V
TP 5 Cable connectors input X1/2 to X1/4 23.3 V A.C. / D.C., tolerance + 1.5 / - 0.3 V
TP 6 Main bulb between cable 8 and 6 max. 22.8 V A.C. / D.C. RMS, tolerance + 0.3 V
Setting of the max. bulb voltage see page 63 - 64, by means of adjustable resistor R 20
1Wiring harness, from electronic to reflector con-
nector,
part no. 65769
6Connection to main bulb, marked with 6
4Connection to reserve bulb, marked with 4
8Common connection to main and reserve bulb,
marked with 8