Service manual

42
CHROMOPHARE® D 650plus, D 530plus, D 500plus, D 650, D 530, D 500 Service Manual (E)
Light intensity power circuit board, part no. 65604
Attention!
No A.C. voltage must be fed into the electronics. A
direct supply of A.C. voltage always destroys the light
electronics.
If only A.C. voltage is available, the filtering board (part
no. 65606) has always to be preconnected to the light
electronics.
If only D.C. voltage is available, preconnecting of the
filtering board is not necessary.
If A.C. voltage (main supply) and D.C. voltage (supple-
mentary supply) are available, the filtering board (part
no. 65606) has always to be preconnected to the light
electronics in the main supply circuitry.
ainput 23 – 26 V D.C.
boutput to light body max. 22,6 V A.C./D.C.
1black - (negative)
2black - (negative)
3red + (positive)
4red + (positive)