Service manual

18
CHROMOPHARE® D 650plus, D 530plus, D 500plus, D 650, D 530, D 500 Service Manual (E)
3.2.2 Adjusting the field size with the centre hand grip
By turning the sterilizable centre hand grip the size of the light field is changed. By turning the hand grip in the
clockwise direction the light field is enlarged, by turning anti-clockwise the light field is reduced. The surgeon must
perform the optimum field setting himself in accordance with the size of the operation opening. There is an opti-
mum contrast when the field size is set approx. 2 - 3 cm larger than the operation opening, so that the edge of the
sterile covering is also illuminated. Thus the adaptation of the surgeon’s eyes in the whole field of vision is im-
proved.
3.2.3Adjusting the field size - light field plotting model LFN (option only D 650plus, D 650)
The CHROMOPHARE
®
D 650plus and D 650 with light field plotting allows the field size to be adjusted by an
electric motor.
By pressing and holding down Key 8 (see control unit page 14) the light field is continuously enlarged up to the
maximum value. By repeated short presses the light field is enlarged in steps.
By pressing and holding down Key 9 (see control unit page 14) the light field is continuously reduced down to the
minimum value. By repeated short presses the light field is reduced in steps.
3.2.4 Changing the sterizable hand grip
By pressing the release button the hand grip sleeve is released and can be re-
moved from the sleeve mounting. The new sterilized replacement sleeve is
pushed onto the sleeve mounting until it touches the stop and is then turned in
a clockwise direction until the sleeve locks into place.
3.2.5Changing the sterizable hand grip light field plotting model LFN (option only D 650plus, D 650)
The sterilizable hand grip sleeve is released by pressure on the bolt on the
upper collar of the sleeve. It can then be removed from the sleeve
mounting.
The new sterilized replacement sleeve is pushed onto the sleeve mounting
until it touches the stop and is then turned in a clockwise direction until
the sleeve is heard locking into place.