Service manual
15
CHROMOPHARE® D 650plus, D 530plus, D 500plus, D 650, D 530, D 500 Service Manual (E)
4By pressing and holding down Key 4 (DARK) the illuminance is continuously reduced down to the minimum
value. Repeated short presses decrease the illuminance in steps.
5If the indicator light shows red, there is a fault in the control unit electronics
6Only with the EndoLite® option. By pressing Key 6 EndoLite
®
the surgical area illumination is switched off and
the surround lighting is switched on
7Only with the EndoLite® option. By pressing Key 7 (OP light) the EndoLite
®
surround lighting is switched off
and the surgical area lighting is switched on again.
8Only with the LFN model. By pressing and holding down Key 8 (field size large) the light field is continuously
enlarged up to the maximum value. Repeated short presses enlarge the light field in steps.
9Only with the LFN model. By pressing and holding down Key 9 (field size small) the light field is continuously
reduced down to the minimum value. Repeated short presses reduce the light field in steps.
Booster activation
1.Press key 4 (Bright) until LED flashes up in the key (max. brightness).
2.The light booster will be activated by pressing simultaneously the key 1 (ON) and 3 (Bright). Luminous power
rises from 110 kLux to 130 kLux.
Note!
The light booster is deactivated by switching off the light with key 2 (OFF).