Operating instructions
6
CHROMOPHARE
®
D 650plus, D 530plus,D 300, C 95x Operating instructions (E)
1.3 General information
• This user’s manual is considered part of the equipment. It must be kept in the vicinity of the equipment at all
times. Precise observance of the user’s manual is a prerequisite for the proper use and correct operation of the
equipment, which is essential for the safety of patients and operators alike.
• Only accessories which are specified in this user’s manual, and which have been tested together with the equip-
ment, may be used. If accessories are used which are not specified in the user’s manual, their ability to be used in
accordance with safety regulations must be proved.
• All literature relates to the equipment model and the prevailing basic safety regulations when printed. All rights
are reserved for equipment, switches, procedures, software programs and names.
1.4 Information about product liability
BERCHTOLD considers itself responsible for the consequences of safety, reliability and performance of the
equipment only if
a) installation, modifications or repairs have been performed only by BERCHTOLD, or by an agent expressly
authorized by BERCHTOLD to do so,
b) the electrical installation of the room complies with regulations IEC 60364-7-10,
c) the equipment is used in accordance with the user’s manual.
1.5 Receiving inspection
Please inspect the equipment and accessories immediately after receipt for any transportation damage and defects.
Claims for damage will only be valid if the vendor (BERCHTOLD) or the carrier are informed
at once. A damage report must then be prepared immediately. The damage record must be submitted to the near-
est BERCHTOLD representative or to BERCHTOLD directly.
When an appliance or parts of an appliance are returned to BERCHTOLD or to a BERCHTOLD service agent,
the original packaging should be used where possible. The following accompanying documentation must be at-
tached: owner’s name and address, equipment number (see type plate), description of the defect.
1.5.1
For transport and storage (in original packing, ; for maximum 15 weeks)
Ambient temperature - between –10°C and +60°C
Relative air moisture - between 10 % and 85 %, no condensation
Air pressure - between 500 hPa and 1060 hPa
For operation
Ambient temperature - between +10°C and +40°C
Relative air moisture - between 30 % and 75 %
Air pressure - between 700 hPa and 1060 hPa
1.6 EC Certification
The equipment complies with the requirements of the EC guideline regarding medical products, 93/42/EEC as
well as the UL guidlines.