Operating instructions
28
CHROMOPHARE
®
D 650plus, D 530plus,D 300, C 95x Operating instructions (E)
6.3.2 Instructions for the changing lamps CHROMOPHARE
®
D 650plus, D 530plus
The symbol with the meaning ‘Danger - Risk of Burning’ is placed on the
lamp support. Do not touch the quartz cone of the halogen bulb with your bare
fingers, use a cloth or soft paper.
Switch off the lighting system and let it cool down (risk of burn-
ing). During replacement of the lamps the patient must not be
touched as it is possible to come into contact with parts under
tension.
Release and remove sterilizable hand grip before.
Unscrew the knurled screw on the underside of the light by
turning anti-clockwise.
Pull out the whole light system at the hand grip (picture 2).
Remove a faulty halogen lamp. We advise users to insert the re-
serve lamp into the main light lamp holder , marked with <R>,
and to insert the new halogen lamp in the reserve lamp holder so
that the reserve lamp does not become too worn.