Operating instructions

26
CHROMOPHARE
®
D 650plus, D 530plus,D 300, C 95x Operating instructions (E)
6.3 Instructions for changing lamps
The halogen bulb has an average life of 1000 hours. In the event
of failure of the main lamp, switch over to a reserve lamp occurs
automatically and the reserve lamp swivels automatically into the
former position of the main lamp.
A yellow indicator light up on the lamp housing
D 650plus/D 530plus and a red indicator light up on the key-
board C 95x when the lamp safety system has switched over to
the reserve lamp. The defective main lamp should then be ex-
changed as quickly as possible.
6.3.1 Instructions for the changing lamps CHROMOPHARE
®
C 95x
The symbol with the meaning ‘Danger - Risk of Burning’ is placed on the
lamp support. Do not touch the quartz cone of the halogen bulb with your bare
fingers, use a cloth or soft paper.
Switch off the lighting system and let it cool down (risk of
burning).
Release and remove sterilizable handle.
Unscrew the knurled screw on the underside of the light by
turning anti-clockwise.
Pull out the whole light system at the hand grip.