4. OPERATION
4.1 Control Panel and symbols CHROMOPHARE
®
C 95x
The following functions can be controlled by the control panel:
1 ON Switch on the light
2 OFF Switch off the light
3 Red indicator LED indicates main light is faulty
4 Field size smal Reduce field size
5 Field size large Enlarge field size
6 Bright Increase intensity
7 Dark Decreases intensity
8 Column indicator Indicates the intensity of luminosity selected
9 <Stopp> key Interrupts AB and LFN.
By pressing the <On> key, AB and LFN are re-enabled
10 Red indicator Warning signal indicating a breakdown in the control circuit
When the <Stop> (9) key is pressed, the indicator will remain illuminated
(ON) The OP light is switched on by pressing Key 1.
(OFF) The OP light is switched off by pressing Key 2.
(Bright) By pressing and holding down Key 6 the illuminance is continuously increased up to the maxi-
mum value. Repeated short presses increase the illuminance in steps.
(Dark) By pressing and holding down Key 7 the illuminance is continuously reduced down to the
minimum value. Repeated short presses decrease the illuminance in steps.
(field size small)By pressing and holding down Key 4 the light field is continuously reduced down to the mini-
mum value. Repeated short presses reduce the light field in steps.
(field size large) By pressing and holding down Key 5 the light field is continuously enlarged up to the maximum
value. Repeated short presses enlarge the light field in steps.